. 2012 Jan;9(1):2-7.
doi: 10.1016/j.hrthm.2011.08.022. Epub 2011 Aug 26.

Novel method of signal analysis for ablation of Wolff-Parkinson-White syndrome

Affiliations

Novel method of signal analysis for ablation of Wolff-Parkinson-White syndrome

Scott R Ceresnak et al. Heart Rhythm. 2012 Jan.

Abstract

Background: Identification of the site of successful radiofrequency catheter ablation (RFCA) for Wolff-Parkinson-White (WPW) syndrome may be subjective.

Objective: The purpose of this study was to develop an automated signal analysis program to predict a successful ablation site.

Methods: Patients who underwent successful RFCA for WPW from 2008- 2010 at our center were analyzed. Inclusion criteria were age <21 years, loss of preexcitation in <5 seconds, and sustained success at 3 months. Exclusion criteria were congenital heart disease and pacing during RFCA. The standard recording system signal was filtered into low frequency (LF 0-≤0.02 Hz) and high frequency (HF >0.02-≤0.45 Hz). Software identified the beginning of the HF signal, LF and R-wave peaks, LF/HF signal amplitude, and area under the HF/LF signals. Successful and unsuccessful (radiofrequency energy applied without accessory pathway block) signals were compared.

Results: Thirty patients were analyzed; 16 had both successful and unsuccessful signals, and 14 had a successful ablation with 1 radiofrequency application. Mean age was 13.7 ± 3.1 years, weight 54.9 ± 22.4 kg, and time to accessory pathway ablation 1.7 ± 1.4 seconds. Significant differences were found between successful and unsuccessful signals in area under HF signal, LF amplitude, LF to R time, HF ratio, and HF area × HF ratio. A receiver operating curve of HF area × HF ratio produced an area under the curve of 0.89. An HF area × HF ratio of 3.1 distinguished successful from unsuccessful signals with 100% specificity and 81% sensitivity.

Conclusion: Automated signal analysis retrospectively differentiated successful from unsuccessful signals in patients undergoing RFCA for WPW. This software may improve the safety and efficacy of RFCA in children.