A comparison of endotracheal bioimpedance cardiography and transpulmonary thermodilution in cardiac surgery patients
- PMID: 21880510
- DOI: 10.1053/j.jvca.2011.06.023
A comparison of endotracheal bioimpedance cardiography and transpulmonary thermodilution in cardiac surgery patients
Abstract
Objectives: The authors hypothesized that bioimpedance cardiography measured by the Endotracheal Cardiac Output Monitor (ECOM; ConMed, Utica, NY) is a convenient and reliable method for both cardiac index (CI) assessment and prediction in fluid responsiveness.
Design: A prospective observational study.
Setting: A teaching university hospital.
Participants: Twenty-five adult patients.
Interventions: Admission to the intensive care unit after conventional cardiac surgery and investigation before and after a fluid challenge.
Measurements and main results: Simultaneous comparative CI data points were collected from transpulmonary thermodilution (TD) and ECOM. Correlations were determined by linear regression. Bland-Altman analysis was used to compare the bias, precision, and limits of agreement. The percentage error was calculated. Pulse-pressure variations (PPVs) and stroke-volume variations (SVVs) before fluid challenge were collected to assess their discrimination in predicting fluid responsiveness. A weak but statistically significant relationship was found between CI(TD) and CI(ECOM) (r = 0.31, p = 0.03). Bias, precision, and limits of agreement between CI(TD) and CI(ECOM) were 0.08 L/min/m(2) (95% confidence interval, -0.11 to 0.27), 0.68 L/min/m(2), and -1.26 to 1.42 L/min/m(2), respectively. The percentage error was 51%. A nonsignificant positive relationship was found between percent changes in CI(TD) and CI(ECOM) after fluid challenge (r = 0.37, p = 0.06). Areas under the ROC curves for both PPV and SVV to predict fluid responsiveness were 0.86 (95% confidence interval, 0.67-1.06) and 0.89 (95% confidence interval, 0.74-1.04, respectively; p = 0.623).
Conclusions: Continuous measurements of CI under dynamic conditions are consistent and easy to obtain with ECOM although not interchangeable with transpulmonary thermodilution. SVV given by ECOM is a dynamic parameter that predicts fluid responsiveness with good accuracy and discrimination.
Copyright © 2012 Elsevier Inc. All rights reserved.
Similar articles
-
Cardiac output determination from endotracheally measured impedance cardiography: clinical evaluation of endotracheal cardiac output monitor.J Cardiothorac Vasc Anesth. 2011 Oct;25(5):770-5. doi: 10.1053/j.jvca.2010.12.017. Epub 2011 Feb 26. J Cardiothorac Vasc Anesth. 2011. PMID: 21354828
-
Assessment of changes in cardiac index and fluid responsiveness: a comparison of Nexfin and transpulmonary thermodilution.Acta Anaesthesiol Scand. 2013 Jul;57(6):704-12. doi: 10.1111/aas.12108. Epub 2013 Mar 25. Acta Anaesthesiol Scand. 2013. PMID: 23521161 Clinical Trial.
-
Cardiac output measurement by bioimpedance and noninvasive pulse contour analysis compared with the continuous pulmonary artery thermodilution technique.J Cardiothorac Vasc Anesth. 2014 Jun;28(3):534-9. doi: 10.1053/j.jvca.2014.01.007. Epub 2014 Apr 16. J Cardiothorac Vasc Anesth. 2014. PMID: 24746334
-
Minimally invasive measurement of cardiac output during surgery and critical care: a meta-analysis of accuracy and precision.Anesthesiology. 2010 Nov;113(5):1220-35. doi: 10.1097/ALN.0b013e3181ee3130. Anesthesiology. 2010. PMID: 20881596 Review.
-
Cardiac Output Measurement in Neonates and Children Using Noninvasive Electrical Bioimpedance Compared With Standard Methods: A Systematic Review and Meta-Analysis.Crit Care Med. 2022 Jan 1;50(1):126-137. doi: 10.1097/CCM.0000000000005144. Crit Care Med. 2022. PMID: 34325447
Cited by
-
The use of pulse pressure variation for predicting impairment of microcirculatory blood flow.Sci Rep. 2021 Apr 28;11(1):9215. doi: 10.1038/s41598-021-88458-3. Sci Rep. 2021. PMID: 33911116 Free PMC article.
-
Early goal-directed therapy based on endotracheal bioimpedance cardiography: a prospective, randomized controlled study in coronary surgery.J Clin Monit Comput. 2015 Jun;29(3):351-8. doi: 10.1007/s10877-014-9611-5. Epub 2014 Nov 9. J Clin Monit Comput. 2015. PMID: 25380955 Clinical Trial.
-
Can endotracheal bioimpedance cardiography assess hemodynamic response to passive leg raising following cardiac surgery?Ann Intensive Care. 2012 Jul 11;2(1):26. doi: 10.1186/2110-5820-2-26. Ann Intensive Care. 2012. PMID: 22784815 Free PMC article.
-
Does pulse pressure variation predict fluid responsiveness in critically ill patients? A systematic review and meta-analysis.Crit Care. 2014 Nov 27;18(6):650. doi: 10.1186/s13054-014-0650-6. Crit Care. 2014. PMID: 25427970 Free PMC article.
-
Applications of minimally invasive cardiac output monitors.Int J Emerg Med. 2012 Apr 24;5:18. doi: 10.1186/1865-1380-5-18. Int J Emerg Med. 2012. PMID: 22531454 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical