Berlin Heart EXCOR Pediatric ventricular assist device Investigational Device Exemption study: study design and rationale
- PMID: 21884857
- DOI: 10.1016/j.ahj.2011.05.026
Berlin Heart EXCOR Pediatric ventricular assist device Investigational Device Exemption study: study design and rationale
Abstract
Background: Currently, there are no Food and Drug Administration-approved devices available that can provide long-term mechanical circulatory support to smaller children with severe heart failure as a bridge to heart transplant (HT). In recent years, the Berlin Heart EXCOR Pediatric ventricular assist device (VAD) has emerged as a potential treatment option. Systematic data on the safety and efficacy of the EXCOR are limited.
Methods: The Investigational Device Exemption (IDE) clinical study is designed to evaluate the safety and probable benefit of the EXCOR to support regulatory review of the device under the Humanitarian Device Exemption regulation. The study design and rationale are reviewed in light of the well-described challenges inherent in small population studies.
Results: The Berlin Heart EXCOR IDE clinical study is a prospective, multicenter, single-arm, clinical cohort study. Children aged 0 to 16 years with severe heart failure (Interagency Registry for Mechanically Assisted Circulatory Support profile 1 or 2) due to 2-ventricle heart disease and actively listed for HT comprise the primary study cohort. The control population is a propensity-matched retrospective cohort of children supported with extracorporeal membrane oxygenation, the only bridge device available to smaller children before the EXCOR. The primary efficacy end point is survival to heart transplantation or recovery. The primary safety end point is the incidence of serious adverse events as defined by pediatric Interagency Registry for Mechanically Assisted Circulatory Support criteria. The study will enroll a total of 48 subjects in 2 cohorts based on body surface area (cohort 1 <0.7 m(2), cohort 2 0.7-1.5 m(2)) and is powered to show safety superiority to a prespecified performance goal of 0.25 serious adverse events per day of support. Children ineligible for the primary cohort will still have access to the device in a third compassionate-use cohort where adverse event data will be collected for additional safety characterization of the device.
Conclusion: The Berlin Heart IDE clinical study will be the first bridge-to-HT VAD study designed exclusively for children. It is anticipated that the study will provide important information on the safety and efficacy of the Berlin Heart EXCOR Pediatric in children while providing valuable lessons into the design and conduct of future VAD studies in children.
Trial registration: ClinicalTrials.gov NCT00583661.
Copyright © 2011 Mosby, Inc. All rights reserved.
Similar articles
-
The use of the Berlin Heart EXCOR in patients with functional single ventricle.J Thorac Cardiovasc Surg. 2014 Feb;147(2):697-704; discussion 704-5. doi: 10.1016/j.jtcvs.2013.10.030. Epub 2013 Nov 28. J Thorac Cardiovasc Surg. 2014. PMID: 24290716
-
Post-transplant outcomes of children bridged to transplant with the Berlin Heart EXCOR Pediatric ventricular assist device.Circulation. 2013 Sep 10;128(11 Suppl 1):S24-31. doi: 10.1161/CIRCULATIONAHA.112.000446. Circulation. 2013. PMID: 24030413
-
Berlin Heart EXCOR pediatric ventricular assist device for bridge to heart transplantation in US children.Circulation. 2013 Apr 23;127(16):1702-11. doi: 10.1161/CIRCULATIONAHA.112.000685. Epub 2013 Mar 28. Circulation. 2013. PMID: 23538380 Clinical Trial.
-
Cardiac extracorporeal life support: state of the art in 2007.Cardiol Young. 2007 Sep;17 Suppl 2:104-15. doi: 10.1017/S1047951107001217. Cardiol Young. 2007. PMID: 18039404 Review.
-
Mechanical circulatory support for infants and small children.Semin Thorac Cardiovasc Surg Pediatr Card Surg Annu. 2011;14(1):38-44. doi: 10.1053/j.pcsu.2011.01.008. Semin Thorac Cardiovasc Surg Pediatr Card Surg Annu. 2011. PMID: 21444048 Review.
Cited by
-
An introduction to point-of-care testing in extracorporeal circulation and LVADs.Hematology Am Soc Hematol Educ Program. 2018 Nov 30;2018(1):516-521. doi: 10.1182/asheducation-2018.1.516. Hematology Am Soc Hematol Educ Program. 2018. PMID: 30504352 Free PMC article. Review.
-
Mechanical circulatory support in pediatrics.Ann Cardiothorac Surg. 2014 Sep;3(5):507-12. doi: 10.3978/j.issn.2225-319X.2014.08.25. Ann Cardiothorac Surg. 2014. PMID: 25452912 Free PMC article.
-
Management of anticoagulation and antiplatelet therapy in patients with left ventricular assist devices.J Thromb Thrombolysis. 2015 Apr;39(3):337-44. doi: 10.1007/s11239-014-1162-6. J Thromb Thrombolysis. 2015. PMID: 25549823
-
Platelet-mapping assay for monitoring antiplatelet therapy during mechanical circulatory support in children: A retrospective observational study.Res Pract Thromb Haemost. 2017 Jun 20;1(1):120-127. doi: 10.1002/rth2.12010. eCollection 2017 Jul. Res Pract Thromb Haemost. 2017. PMID: 30046680 Free PMC article.
-
Enlarged and echogenic kidneys while on a pediatric ventricular assist device.J Extra Corpor Technol. 2013 Dec;45(4):248-50. J Extra Corpor Technol. 2013. PMID: 24649573 Free PMC article.
Publication types
MeSH terms
Associated data
Grants and funding
LinkOut - more resources
Full Text Sources
Medical