. 2012 Mar;23(3):335-40.
doi: 10.1007/s00192-011-1554-5. Epub 2011 Sep 2.

Anterior sacrospinous ligament fixation associated with paravaginal repair using the Pinnacle device: an anatomical study

Affiliations

Anterior sacrospinous ligament fixation associated with paravaginal repair using the Pinnacle device: an anatomical study

Mélanie Cayrac et al. Int Urogynecol J. 2012 Mar.

Abstract

Introduction and hypothesis: The objective of this paper is to study the reproducibility and anatomical risks of anterior sacrospinous ligament (SSL) fixation associated with paravaginal repair using the Pinnacle device (Boston Scientific).

Methods: Simplified bilateral anterior SSL fixation associated with paravaginal fixation through the arcus tendineus fascia pelvis (ATFP) was performed on five fresh cadavers using the Pinnacle device. Cadaver dissection was then performed by open pelvic surgery.

Results: Eight SSL and ten ATFP were available for analysis. SSL fixations were optimal in four cases, too superficial in three cases, and too high in one case. Mean distance between SSL fixation and ischial spine was 18.6 mm (range 10 to 30 mm). Mean distance between SSL fixation and pudendal nerve was 6.5 mm (range 0 to 15 mm). ATFP fixations were optimal in five cases, good in four cases, and too superficial in one case. In one case (10%), the middle arm of the prosthesis was in contact with the ureter and traction on that arm resulted in ureteral kinking.

Conclusions: Anterior SSL fixation associated with paravaginal repair using the Pinnacle™ device was not reproducible every time in this cadaver study. These results confirm the need for specific training before starting anterior SSL fixations.

PubMed Disclaimer

References

    1. Int Urogynecol J Pelvic Floor Dysfunct. 2009 Mar;20(3):363-5 - PubMed
    1. Prog Urol. 2009 Dec;19(13):907-15 - PubMed
    1. Prog Urol. 2008 Dec;18(13):1068-74 - PubMed
    1. Eur J Obstet Gynecol Reprod Biol. 2002 Feb 10;101(1):74-8 - PubMed
    1. BJOG. 2007 May;114(5):522-31 - PubMed