. 2011;5(Suppl 1):46-9.
doi: 10.2174/1874431101105010046. Epub 2011 Jul 27.

Application study of vascular interventional robotic mechanism for remote steering

Affiliations

Application study of vascular interventional robotic mechanism for remote steering

Zengmin Tian et al. Open Med Inform J. 2011.

Abstract

Background: Recently, robotic systems have been introduced as a useful method for surgical procedures. But in the field of vascular interventional therapy, the development of robotic system is slower.

Objective: The purpose of the study is to verify the reliability and safety of vascular interventional robotic system used in angiography, by the way of in vitro preliminary experiments and animal experiments.

Method: The approach is to employ a proprietary vascular interventional robot system to complete glass vessel models and animal angiogram experiments. This robot system consists of a console port (remote steering system), an assistant port (propelled and rotation system) and a hydraulic fixing device, upon which surgeons control remotely to make go forward and rotate in the glass vessel models and animal vessels, on the 3D operation interface. Consequently, the operation time and success rate are counted and evaluated.

Result: In the glass vessel model experiments, the Catheter can enter various kinds of vessel models with inside diameter length greater than 3mm and angle less than 90(o). In the animal (adult dogs) experiments, surgeons can accomplish smoothly the angiogram of the renal artery, the vertebral renal and the arteria carotis communis, without any complications of surgery.

Conclusion: The angiogram by using vascular interventional robot system is safe and reliable. Surgeons can finish the angiogram in part by remote operation, and the result of angiogram can meet a number of simple expectations. However without wire control and force feedback systems, the applicability of this kind of robot system is not flexible enough and need to be improved in the future.

Keywords: Robotic; animal experiments.; remote steering; vascular interventional.

PubMed Disclaimer

Figures

Fig. (1)
Fig. (1)
Vascular model experiment.
Fig. (2)
Fig. (2)
Animal experiment.

Similar articles

References

    1. Mack MJ. Minimally invasive and robotic surgery. JAMA. 2001;285:568–72. - PubMed
    1. Da l Zhang D, Wang T. Overview of the vascular interventional robot. Int J Med Robot. 2008;4:289–94. - PubMed
    1. Saliba W, Cummings JE, Oh S, et al. Novel robotic catheter remote control system: feasibility and safety of transseptal puncture and endocardial catheter navigation. J Cardiovasc Electrophysiol. 2006;17:1102–5. - PubMed
    1. Riga CV, Bicknell CD, Wallace D, et al. Robot-assisted antegrade in-situ fenestrated stent grafting. Cardiovasc Intervent Radiol. 2009;32:522–4. - PubMed
    1. Advanced surgical service robotics in the European Union 6th and 7th Framework Programs, 2010 IEEE International Conference on Robotics and Automation, 2010; Anchorage, Alaska.

LinkOut - more resources