Non-invasive measurement of local pulse pressure by pulse wave-based ultrasound manometry (PWUM)
- PMID: 21904023
- PMCID: PMC4005495
- DOI: 10.1088/0967-3334/32/10/012
Non-invasive measurement of local pulse pressure by pulse wave-based ultrasound manometry (PWUM)
Abstract
Central blood pressure (CBP) has been established as a relevant indicator of cardiovascular disease. Despite its significance, CBP remains particularly challenging to measure in standard clinical practice. The objective of this study is to introduce pulse wave-based ultrasound manometry (PWUM) as a simple-to-use, non-invasive ultrasound-based method for quantitative measurement of the central pulse pressure. Arterial wall displacements are estimated using radiofrequency ultrasound signals acquired at high frame rates and the pulse pressure waveform is estimated using both the distension waveform and the local pulse wave velocity. The method was tested on the abdominal aorta of 11 healthy subjects (age 35.7 ± 16 y.o.). PWUM pulse pressure measurements were compared to those obtained by radial applanation tonometry using a commercial system. The average intra-subject variability of the pulse pressure amplitude was found to be equal to 4.2 mmHg, demonstrating good reproducibility of the method. Excellent correlation was found between the waveforms obtained by PWUM and those obtained by tonometry in all subjects (0.94 < r < 0.98). A significant bias of 4.7 mmHg was found between PWUM and tonometry. PWUM is a highly translational method that can be easily integrated in clinical ultrasound imaging systems. It provides an estimate of the pulse pressure waveform at the imaged location, and may offer therefore the possibility to estimate the pulse pressure at different arterial sites. Future developments include the validation of the method against invasive estimates on patients, as well as its application to other large arteries.
Figures




References
-
- World Health Organization-International Society of Hypertension Guidelines for the Management of Hypertension. Guidelines Subcommittee. J. Hypertens. 1999;17:151–183. - PubMed
-
- Avolio A, Van Bortel L, Boutouyrie P, Cockcroft J, McEniery C, Protogerou A, Roman M, Safar M, Segers P, Smulyan H. Role of pulse pressure amplification in arterial hypertension: Experts' opinion and review of the data. Hypertension. 2009;54:375–383. - PubMed
-
- Beulen BW, Bijnens N, Koutsouridis GG, Brands PJ, Rutten MC, van de Vosse FN. Toward noninvasive blood pressure assessment in artieries by using ultrasound. Ultrasound Med. Biol. 2011;37:788–797. - PubMed
-
- Brands PJ, Willigers JM, Ledoux LAF, Reneman RS, Hoeks APG. A noninvasive method to estimate pulse wave velocity in arteries locally by means of ultrasound. Ultrasound Med. Biol. 1998;24:1325–1335. - PubMed
-
- Chen C-H, Nevo E, Fetics B, Pak P, Yin F, Maughan W, Kass D. Estimation of Central aortic pressure waveform by mathematical transformation of radial tonometry pressure: Validation of generalized transfer function. Circulation. 1997;95:1827–1836. - PubMed
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources
Other Literature Sources