Posterior capsule opacification and capsular bag performance of a microincision intraocular lens

Affiliations
Randomized Controlled Trial

Posterior capsule opacification and capsular bag performance of a microincision intraocular lens

Vinod Gangwani et al. J Cataract Refract Surg. 2011 Nov.

Abstract

Purpose: To assess the axial stability and posterior capsule opacification (PCO)-inhibiting efficacy of a microincision hydrophilic intraocular lens (IOL) (Idea 613 XC) and a standard hydrophobic open-loop acrylic IOL (Acrysof SA60AT).

Setting: Moorfields Eye Hospital NHS Foundation Trust, London, United Kingdom.

Design: Clinical trial.

Methods: Patients with bilateral cataract received a microincision cataract surgery (MICS) IOL in 1 eye and a small-incision cataract surgery (SICS) IOL in the contralateral eye. Follow-up examinations were performed immediately after surgery and postoperatively at 1 week and 3 and 24 months. Posterior capsule opacification was assessed using retroillumination photography. Anterior chamber depth (ACD) was measured using partial coherence laser interferometry, and IOL tilt and decentration were measured using a Purkinje meter.

Results: Seventy eyes (35 patients) were enrolled. The mean ACD at 3 months was 4.91 mm ± 0.26 (SD) in the MICS group and 4.60 ± 0.23 mm in the SICS group. The anterior IOL shift over the first 3 months was slightly smaller with the MICS IOL (81 μm) than with the SICS IOL (254 μm). The mean ACD was 0.377 ± 178 μm deeper with the MICS IOL (P<.01, paired t test). Two years postoperatively, the amount of regeneratory PCO was higher with the MICS IOL than with the SICS IOL (image analysis software score 2.6 versus 1.9; P=.005).

Conclusions: Two years after surgery, the PCO rate was slightly higher with the MICS IOL. The 2 IOLs had similar performance in terms of axial stability, decentration, and tilt.

Financial disclosure: No author has a financial or proprietary interest in any material or method mentioned. Additional disclosures are found in the footnotes.