Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Review
. 1990 Mar:84:57-73.
doi: 10.1289/ehp.908457.

Calcium measurements with electron probe X-ray and electron energy loss analysis

Affiliations
Review

Calcium measurements with electron probe X-ray and electron energy loss analysis

A LeFurgey et al. Environ Health Perspect. 1990 Mar.

Abstract

This paper presents a broad survey of the rationale for electron probe X-ray microanalysis (EPXMA) and the various methods for obtaining qualitative and quantitative information on the distribution and amount of elements, particularly calcium, in cryopreserved cells and tissues. Essential in an introductory consideration of microanalysis in biological cryosections is the physical basis for the instrumentation, fundamentals of X-ray spectrometry, and various analytical modes such as static probing and X-ray imaging. Some common artifacts are beam damage and contamination. Inherent pitfalls of energy dispersive X-ray systems include Si escape peaks, doublets, background, and detector calibration shifts. Quantitative calcium analysis of thin cryosections is carried out in real time using a multiple least squares fitting program on filtered X-ray spectra and normalizing the calcium peak to a portion of the continuum. Recent work includes the development of an X-ray imaging system where quantitative data can be retrieved off-line. The minimum detectable concentration of calcium in biological cryosections is approximately 300 mumole kg dry weight with a spatial resolution of approximately 100 A. The application of electron energy loss (EELS) techniques to the detection of calcium offers the potential for greater sensitivity and spatial resolution in measurement and imaging. Determination of mass thickness with EELS can facilitate accurate calculation of wet weight concentrations from frozen hydrated and freeze-dried specimens. Calcium has multiple effects on cell metabolism, membrane transport and permeability and, thus, on overall cell physiology or pathophysiology. Cells can be rapidly frozen for EPXMA during basal or altered functional conditions to delineate the location and amount of calcium within cells and the changes in location and concentration of cations or anions accompanying calcium redistribution. Recent experiments in our laboratory document that EPXMA in combination with other biochemical and electrophysiological techniques can be used to study, for example, sodium and calcium compartmentation in cultured cardiac cells. Such analyses can also be used to clarify the role of calcium in anoxic renal cell injury and to evaluate proposed ionic defects in cells of individuals with cystic fibrosis.

PubMed Disclaimer

References

    1. Ultramicroscopy. 1976 Sep-Oct;1(4):317-39 - PubMed
    1. Ann N Y Acad Sci. 1986;483:241-4 - PubMed
    1. Ann N Y Acad Sci. 1986;483:295-310 - PubMed
    1. Ann N Y Acad Sci. 1986;483:311-25 - PubMed
    1. Ann N Y Acad Sci. 1986;483:339-53 - PubMed