Evaluation of a novel normal database with matched SPECT systems and optimal pre-filter parameters for 3D-SSP
- PMID: 21909693
- DOI: 10.1007/s12149-011-0534-7
Evaluation of a novel normal database with matched SPECT systems and optimal pre-filter parameters for 3D-SSP
Abstract
Purpose: A normal perfusion database (NDB) is imperative for the statistical imaging of brain function. This study validates a novel NDB created under the same injection dose and acquisition conditions for three gamma camera systems and evaluates optimal pre-filter parameters for three-dimensional stereotactic surface projections (3D-SSPs).
Method: We compared a novel NDB that matched the databases in each of three vendor gamma camera systems with a conventionally constructed NDB (conventional NDB) and a NDB constructed in-house for 3D-SSP. We generated hypoperfused regions where pre-specified volumes were simulated for various areas in SPECT images. The properties of each NDB were evaluated based on the distribution of the standard deviation (SD). Abnormal accumulation regions were validated using Z, extent, and artifactual scores. Detection error was used to evaluate the optimal Butterworth pre-filter cutoff frequency with the perfusion defect rate (PDR) in 3D-SSP.
Results: The SD distribution was the same in the novel NDB and in the NDB constructed in-house, and the SD of the peak distribution was 0.08-0.07. The Z and extent scores of the novel DB and the NDB constructed in-house were similar, but increased along with the artifactual scores when using the conventional NDB. Many artifacts appeared in the Z score map when using the conventional NDB. The detection error deviated from the actual value by -1.3% at a cutoff frequency of 0.58 cycles/cm and a PDR of 30%, which was the lowest. The cutoff frequency became lower or higher, and the low-perfusion defect rate increased according to the increasing detection error. The optimal cutoff frequency was between 0.52 and 0.58 cycles/cm.
Conclusions: We generated a novel NDB according to the individual devices and compared it with a conventional and a NDB constructed in-house. The Z and extent scores were essentially equal when using the novel DB and the NDB constructed in-house, but considerably differed when using the conventional NDB. The optimal cutoff frequency of the Butterworth filter evaluated from the detection error was in the range of 0.52-0.58 cycles/cm. The detection error increased the perfusion defect rate by <15% and this was undetectable in 3D-SSP. The next step will be to improve the accuracy of the extent of abnormal regions and the sensitivity of the Z score using a novel NDB constructed according to the individual devices.
Similar articles
-
Does applying resolution recovery to normal databases confer an advantage over conventional 3D-stereotactic surface projection techniques?Radiol Phys Technol. 2017 Jun;10(2):240-248. doi: 10.1007/s12194-017-0391-8. Epub 2017 Feb 4. Radiol Phys Technol. 2017. PMID: 28161808
-
Impact of injected dose and acquisition time on a normal database by use of 3D-SSP in SPECT images: quantitative simulation studies.Radiol Phys Technol. 2015 Jul;8(2):224-31. doi: 10.1007/s12194-015-0311-8. Epub 2015 Mar 11. Radiol Phys Technol. 2015. PMID: 25758210
-
Effect of prefiltering cutoff frequency and scatter and attenuation corrections during normal database creation for statistical imaging analysis of the brain.J Nucl Med Technol. 2011 Sep;39(3):231-6. doi: 10.2967/jnmt.110.079871. Epub 2011 Jul 27. J Nucl Med Technol. 2011. PMID: 21795375
-
[Neurological diseases and SPECT--analysis using easy Z-score imaging system (eZIS)].Brain Nerve. 2007 May;59(5):487-93. Brain Nerve. 2007. PMID: 17533974 Review. Japanese.
-
Stereotactic statistical imaging analysis of the brain using the easy Z-score imaging system for sharing a normal database.Radiat Med. 2006 Aug;24(7):545-52. doi: 10.1007/s11604-006-0056-8. Radiat Med. 2006. PMID: 17058152 Review.
Cited by
-
Computer-assisted system for diagnosing degenerative dementia using cerebral blood flow SPECT and 3D-SSP: a multicenter study.Jpn J Radiol. 2014 Jul;32(7):383-90. doi: 10.1007/s11604-014-0329-6. Epub 2014 May 17. Jpn J Radiol. 2014. PMID: 24838777
-
Does applying resolution recovery to normal databases confer an advantage over conventional 3D-stereotactic surface projection techniques?Radiol Phys Technol. 2017 Jun;10(2):240-248. doi: 10.1007/s12194-017-0391-8. Epub 2017 Feb 4. Radiol Phys Technol. 2017. PMID: 28161808
-
Impact of injected dose and acquisition time on a normal database by use of 3D-SSP in SPECT images: quantitative simulation studies.Radiol Phys Technol. 2015 Jul;8(2):224-31. doi: 10.1007/s12194-015-0311-8. Epub 2015 Mar 11. Radiol Phys Technol. 2015. PMID: 25758210