Increasing density of wireless communication and development of radio frequency identification (RFID) technology in particular have increased the susceptibility of patients equipped with cardiac rhythmic monitoring devices (CRMD) to environmental electro magnetic interference (EMI). Several organizations reported observing CRMD EMI from different sources. This paper focuses on mathematically analyzing the energy as perceived by the implanted device, i.e., voltage. Radio frequency (RF) energy transmitted by RFID interrogators is considered as an example. A simplified front-end equivalent circuit of a CRMD sensing circuitry is proposed for the analysis following extensive black-box testing of several commercial pacemakers and implantable defibrillators. After careful understanding of the mechanics of the CRMD signal processing in identifying the QRS complex of the heart-beat, a mitigation technique is proposed. The mitigation methodology introduced in this paper is logical in approach, simple to implement and is therefore applicable to all wireless communication protocols.
Similar articles
-
Implantable rhythm devices and electromagnetic interference: myth or reality?Expert Rev Cardiovasc Ther. 2008 Jul;6(6):823-32. doi: 10.1586/14779072.6.6.823. Expert Rev Cardiovasc Ther. 2008. PMID: 18570620 Review.
-
In vitro tests reveal sample radiofrequency identification readers inducing clinically significant electromagnetic interference to implantable pacemakers and implantable cardioverter-defibrillators.Heart Rhythm. 2010 Jan;7(1):99-107. doi: 10.1016/j.hrthm.2009.09.071. Epub 2009 Oct 12. Heart Rhythm. 2010. PMID: 20129290
-
Electromagnetic interference with cardiac pacemakers and implantable cardioverter-defibrillators from low-frequency electromagnetic fields in vivo.Europace. 2013 Mar;15(3):388-94. doi: 10.1093/europace/eus345. Epub 2012 Nov 1. Europace. 2013. PMID: 23125355
-
A call for safer utilization of radio frequency identification in the e-health era.Telemed J E Health. 2011 Oct;17(8):615-9. doi: 10.1089/tmj.2011.0012. Epub 2011 Jul 22. Telemed J E Health. 2011. PMID: 21780943
-
Effects of electromagnetic interference on implanted cardiac devices and their management.Cardiol Rev. 2007 Nov-Dec;15(6):304-9. doi: 10.1097/CRD.0b013e31813e0ba9. Cardiol Rev. 2007. PMID: 18090066 Review.
Cited by
-
Adhoc electromagnetic compatibility testing of non-implantable medical devices and radio frequency identification.Biomed Eng Online. 2013 Jul 11;12:71. doi: 10.1186/1475-925X-12-71. Biomed Eng Online. 2013. PMID: 23845013 Free PMC article.
-
Feasibility results of an electromagnetic compatibility test protocol to evaluate medical devices to radio frequency identification exposure.Biomed Eng Online. 2014 Aug 3;13:110. doi: 10.1186/1475-925X-13-110. Biomed Eng Online. 2014. PMID: 25086451 Free PMC article.
-
Innovation and translation efforts in wireless medical connectivity, telemedicine and eMedicine: a story from the RFID Center of Excellence at the University of Pittsburgh.Ann Biomed Eng. 2013 Sep;41(9):1913-25. doi: 10.1007/s10439-013-0873-8. Epub 2013 Jul 30. Ann Biomed Eng. 2013. PMID: 23897048 Free PMC article.
-
Sensor fusion methods for reducing false alarms in heart rate monitoring.J Clin Monit Comput. 2016 Dec;30(6):859-867. doi: 10.1007/s10877-015-9786-4. Epub 2015 Oct 6. J Clin Monit Comput. 2016. PMID: 26439831
MeSH terms
LinkOut - more resources
Full Text Sources
Medical