Accuracy of the ultrasonic cardiac output monitor in healthy term neonates during postnatal circulatory adaptation
- PMID: 21933558
Accuracy of the ultrasonic cardiac output monitor in healthy term neonates during postnatal circulatory adaptation
Abstract
Background: Echocardiography is regarded as a gold standard for measuring hemodynamic values. The ultrasonic cardiac output monitor (USCOM) is a new method for measuring hemodynamics and could provide non-invasive point of care guidance. So far, there are no published USCOM reference values for neonates, nor has USCOM's accuracy been established in this population. We aimed to determine the accuracy and clinical utility of the USCOM in healthy neonates relative to published echocardiographic data, to establish normal hemodynamic parameters that it measures, and to assess the possible role of USCOM as an alternative to echocardiography as a trend monitor.
Methods: Right and left heart hemodynamics of 90 normal neonates were measured during circulatory adaptation over the first three days of life using the USCOM and automated oscillotonometry.
Results: Heart rate showed a significant decline from days one to three, from 126 to 120 (P < 0.001). Systolic, diastolic and mean arterial pressures all increased significantly from 66 to 71 mmHg, 33 to 38 mmHg and 44 to 49 mmHg, respectively (P < 0.001 in each case). Right ventricular cardiac index (RV-CI) showed no change with a mean of 5.07 L × min(-1) × m(-2). Left ventricular cardiac index (LV-CI) declined from 3.43 to 3.00 L × min(-1) × m(-2) (P < 0.001). RV-CI exceeded LV-CI on all three days by a mean of 61%. The systemic vascular resistance index (SVRI), based on LV-CI, increased significantly over the three days from 1083 to 1403 dyne × sec × cm(-5) × m(2) (P < 0.001).
Conclusions: Normal neonatal hemodynamic values, as indicated by USCOM, were established. LV-CI measurement showed excellent agreement with published echocardiographic studies. RV-CI was constant and exceeded LV-CI for all three days of this study. It may be falsely high due to flow velocity measurement errors arising from the pulmonary branch arteries, and may represent a limitation of the USCOM method. The progressive rise of arterial pressure and SVRI despite a declining LV-CI may indicate functional closure of the ductus arteriosus, with the greatest change occurring within the first 24 hours. Evidence of closure of the foramen ovale was not observed.
Similar articles
-
[Clinical study of neonatal cardiac output measurement methods].Zhonghua Er Ke Za Zhi. 2013 Jan;51(1):58-63. Zhonghua Er Ke Za Zhi. 2013. PMID: 23527933 Chinese.
-
Measurement of systemic oxygen delivery and inotropy in healthy term neonates with the Ultrasonic Cardiac Output Monitor (USCOM).Early Hum Dev. 2013 May;89(5):289-94. doi: 10.1016/j.earlhumdev.2012.10.006. Epub 2012 Nov 17. Early Hum Dev. 2013. PMID: 23164929
-
Nurse-determined assessment of cardiac output. Comparing a non-invasive cardiac output device and pulmonary artery catheter: a prospective observational study.Int J Nurs Stud. 2009 Oct;46(10):1291-7. doi: 10.1016/j.ijnurstu.2009.03.013. Epub 2009 May 6. Int J Nurs Stud. 2009. PMID: 19423107
-
Early and medium term results of tailored therapy for heart failure.Rev Port Cardiol. 2001 Mar;20(3):261-82. Rev Port Cardiol. 2001. PMID: 11417309 Review. English, Portuguese.
-
Ultrasonic evaluation of the heart.Curr Opin Crit Care. 2014 Jun;20(3):309-14. doi: 10.1097/MCC.0000000000000094. Curr Opin Crit Care. 2014. PMID: 24722058 Review.
Cited by
-
[Application of ultrasonic cardiac output monitor in evaluation of cardiac function in children with severe pneumonia].Zhongguo Dang Dai Er Ke Za Zhi. 2016 Sep;18(9):817-820. doi: 10.7499/j.issn.1008-8830.2016.09.006. Zhongguo Dang Dai Er Ke Za Zhi. 2016. PMID: 27655536 Free PMC article. Chinese.
-
Doppler Ultrasound for Heart Rate Assessment in a Porcine Model of Neonatal Asphyxia.Front Pediatr. 2020 Jan 31;8:18. doi: 10.3389/fped.2020.00018. eCollection 2020. Front Pediatr. 2020. PMID: 32083045 Free PMC article.
-
Age-specific non-invasive transcutaneous Doppler ultrasound derived haemodynamic reference ranges in elderly Chinese adults.BBA Clin. 2014 Sep 28;2:48-55. doi: 10.1016/j.bbacli.2014.09.002. eCollection 2014 Dec. BBA Clin. 2014. PMID: 26672648 Free PMC article.
-
Understanding the Haemodynamics of Hypertension.Curr Hypertens Rep. 2018 Apr 10;20(4):29. doi: 10.1007/s11906-018-0832-8. Curr Hypertens Rep. 2018. PMID: 29637390 Review.
-
An Unsettled Promise: The Newborn Piglet Model of Neonatal Acute Respiratory Distress Syndrome (NARDS). Physiologic Data and Systematic Review.Front Physiol. 2019 Oct 30;10:1345. doi: 10.3389/fphys.2019.01345. eCollection 2019. Front Physiol. 2019. PMID: 31736777 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Miscellaneous