Monopolar electrosurgical thermal management for minimizing tissue damage
- PMID: 21947515
- DOI: 10.1109/TBME.2011.2168956
Monopolar electrosurgical thermal management for minimizing tissue damage
Abstract
In this study, a novel thermal management system (TMS) is developed for the minimization of thermal spread created by a monopolar electrosurgical device, the most commonly used surgical instrument. The phenomenon of resistive heating of tissue is modeled using the finite-element method (FEM) to analyze the electrical potential and temperature distributions in biological tissue subjected to heat generation during monopolar electrosurgery. Ex vivo experiments are used to validate the FEM by comparing the model predicted and experimentally measured temperatures. The predicted FEM maximum temperature 1.0 m adjacent to the electrode is within 1% of the experimentally measured maximum temperature using a standard monopolar pencil electrode. A TMS consisting of adjacent cooling channels produces coagulation volumes 80% that of standard monopolar procedures while maintaining comparable temperatures in the targeted tissue below the electrode. In vivo temperatures using a device incorporating a TMS at distances of 2 and 3 m adjacent to the electrode edge are maintained below temperatures known to damage tissue.
© 2011 IEEE
Similar articles
-
Electrosurgical vessel sealing tissue temperature: experimental measurement and finite element modeling.IEEE Trans Biomed Eng. 2013 Feb;60(2):453-60. doi: 10.1109/TBME.2012.2228265. Epub 2012 Nov 20. IEEE Trans Biomed Eng. 2013. PMID: 23192471
-
Thermal effects of the anovo® instrument arms hook electrode and curved scissors: an ex vivo study of safety and precision in monopolar electrosurgery.BMC Surg. 2025 May 29;25(1):236. doi: 10.1186/s12893-025-02969-8. BMC Surg. 2025. PMID: 40442741 Free PMC article.
-
The role of thermal feedback in electrosurgical tissue heating.Technol Health Care. 1995 Oct;3(2):111-6. Technol Health Care. 1995. PMID: 8574761
-
Electrosurgery: part II. Technology, applications, and safety of electrosurgical devices.J Am Acad Dermatol. 2014 Apr;70(4):607.e1-607.e12. doi: 10.1016/j.jaad.2013.09.055. J Am Acad Dermatol. 2014. PMID: 24629362 Review.
-
Active electrode monitoring. How to prevent unintentional thermal injury associated with monopolar electrosurgery at laparoscopy.Surg Endosc. 1998 Aug;12(8):1009-12. doi: 10.1007/s004649900769. Surg Endosc. 1998. PMID: 9685531 Review.
Cited by
-
A continuum thermomechanical model of in vivo electrosurgical heating of hydrated soft biological tissues.Int J Heat Mass Transf. 2018 Dec;127(Pt A):961-974. doi: 10.1016/j.ijheatmasstransfer.2018.07.006. Epub 2018 Jul 14. Int J Heat Mass Transf. 2018. PMID: 30739950 Free PMC article.
-
Local thermal effect of power-on setting on monopolar coagulation: a three-dimensional electrothermal coupled finite element study.Med Biol Eng Comput. 2022 Dec;60(12):3525-3538. doi: 10.1007/s11517-022-02689-8. Epub 2022 Oct 14. Med Biol Eng Comput. 2022. PMID: 36229717
-
A numerical study of the relevance of the electrode-tissue contact area in the application of soft coagulation.Sci Rep. 2025 Jul 31;15(1):27926. doi: 10.1038/s41598-025-13836-0. Sci Rep. 2025. PMID: 40745470 Free PMC article.
-
Energy Dissipation in Ex Vivo Porcine Liver During Electrosurgery.IEEE Trans Biomed Eng. 2017 Jun;64(6):1211-1217. doi: 10.1109/TBME.2016.2595525. Epub 2016 Jul 27. IEEE Trans Biomed Eng. 2017. PMID: 27479955 Free PMC article.
-
Effect of high-frequency electric field on the tissue sticking of minimally invasive electrosurgical devices.R Soc Open Sci. 2018 Jul 11;5(7):180125. doi: 10.1098/rsos.180125. eCollection 2018 Jul. R Soc Open Sci. 2018. PMID: 30109069 Free PMC article.