Comparative Study
. 2011 Dec;42(6):797-802.
doi: 10.1016/j.ejvs.2011.09.003. Epub 2011 Oct 1.

Single-centre prospective comparison between contrast-enhanced ultrasound and computed tomography angiography after EVAR

Affiliations
Free article
Comparative Study

Single-centre prospective comparison between contrast-enhanced ultrasound and computed tomography angiography after EVAR

P Perini et al. Eur J Vasc Endovasc Surg. 2011 Dec.
Free article

Abstract

Aim: To evaluate contrast-enhanced ultrasound (CEUS) as an effective alternative to CT-angiography (CTA) for endoleak detection and aneurismal sac diameter measurement in the follow-up after endovascular abdominal aortic aneurysm repair (EVAR).

Methods: From January 2006 to December 2010, 395 patients underwent EVAR follow-up with both CTA and CEUS. The diameter of the aneurismal sac and the presence of endoleaks were evaluated in all the 395 paired examinations.

Results: Bland-Altman plots showed a good agreement in aneurismal sac diameter evaluation between the two imaging modalities. The mean diameter was 54.93 mm (standard deviation (SD) ±12.57) with CEUS and 56.01 mm (SD ± 13.23) with CTA. The mean difference in aneurismal sac diameter was -1.08 mm ± 3.3543 (95% confidence interval (CI), -0.75 to -1.41), in favour of CTA. The number of observed agreement in endoleak detection was 359/395 (90.89%). The two modalities detected the same type I and type III endoleaks. McNemar's χ(2) test confirmed that CTA and CEUS are equivalent in endoleak detection.

Conclusions: CEUS demonstrated to be as accurate as CTA in endoleak detection and abdominal aortic aneurysm diameter measurements during EVAR follow-up, without carrying the risks of radiation exposure or nephrotoxicity. Even if it cannot be proposed as the sole imaging modality during follow-up, our analysis suggests that it should have a major role.