Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Review
. 2012 Jan;58(1):54-61.
doi: 10.1373/clinchem.2011.165795. Epub 2011 Sep 30.

Analytical characteristics of high-sensitivity cardiac troponin assays

Collaborators, Affiliations
Review

Analytical characteristics of high-sensitivity cardiac troponin assays

Fred S Apple et al. Clin Chem. 2012 Jan.

Erratum in

  • Clin Chem. 2012 Apr;58(4):796

Abstract

Background: Cardiac troponins I (cTnI) and T (cTnT) have received international endorsement as the standard biomarkers for detection of myocardial injury, for risk stratification in patients suspected of acute coronary syndrome, and for the diagnosis of myocardial infarction. An evidence-based clinical database is growing rapidly for high-sensitivity (hs) troponin assays. Thus, clarifications of the analytical principles for the immunoassays used in clinical practice are important.

Content: The purpose of this mini-review is (a) to provide a background for the biochemistry of cTnT and cTnI and (b) to address the following analytical questions for both hs cTnI and cTnT assays: (i) How does an assay become designated hs? (ii) How does one realistically define healthy (normal) reference populations for determining the 99th percentile? (iii) What is the usual biological variation of these analytes? (iv) What assay imprecision characteristics are acceptable? (v) Will standardization of cardiac troponin assays be attainable?

Summary: This review raises important points regarding cTnI and cTnT assays and their reference limits and specifically addresses hs assays used to measure low concentrations (nanograms per liter or picograms per milliliter). Recommendations are made to help clarify the nomenclature. The review also identifies further challenges for the evolving science of cardiac troponin measurement. It is hoped that with the introduction of these concepts, both laboratorians and clinicians can develop a more unified view of how these assays are used worldwide in clinical practice.