. 2011 Oct;6(4):217-20.
doi: 10.4103/1817-1737.84776.

Evaluation of a transcutaneous carbon dioxide monitor in patients with acute respiratory failure

Affiliations

Evaluation of a transcutaneous carbon dioxide monitor in patients with acute respiratory failure

Antonello Nicolini et al. Ann Thorac Med. 2011 Oct.

Abstract

Background: Non-invasive measurement of oxygenation is a routine procedure in clinical practice, but transcutaneous monitoring of PCO(2)(PtCO(2)) is used much less than expected.

Methods: The aim of our study was to analyze the value of a commercially available combined SpO(2)/PtCO(2) monitor (TOSCA-Linde Medical System, Basel, Switzerland) in adult non-invasive ventilated patients with acute respiratory failure. Eighty critically ill adult patients, requiring arterial blood sample gas analyses, underwent SpO(2) and PtCO(2) measurements (10 min after the probe was attached to an earlobe) simultaneously with arterial blood sampling. The level of agreement between PaCO(2) - PtCO(2) and SaO(2) - SpO(2)was assessed by Bland-Altman analyses.

Results: Both, SaO(2) from blood gas analysis and SpO(2) from the transcutaneous monitor, and PaCO(2) and PtCO(2) were equally useful. No measurements were outside of the acceptable clinical range of agreement of ± 7.5 mmHg.

Conclusions: The accuracy of estimation of the TOSCA transcutaneous electrode (compared with the "gold standard" blood sample gas analysis) was generally good. Moreover, TOSCA presents the advantage of the possibility of continuous non-invasive measurement. The level of agreement of the two methods of measurement allows us to state that the TOSCA sensor is useful in routine monitoring of adults admitted to an intermediate respiratory unit and undergoing non-invasive ventilation.

Keywords: Non-invasive measurement; PaCO2; PtCO2; SaO2; SpO2; respiratory failure; transcutaneous electrode.

PubMed Disclaimer

Conflict of interest statement

Conflict of Interest: None declared.

Figures

Figure 1
Figure 1
Bland-Altman plot of difference between two estimation of (a) PO2 and their average (b) PCO2 and their average
Figure 2
Figure 2
SatO2 and PCO2: Averages and 95% confidence limits

Similar articles

Cited by

References

    1. Hutchinson DC, Rocca G, Honeybourne D. Estimation of arterial oxygen tension in adult subjects using transcutaneous electrode. Thorax. 1981;36:473–7. - PMC - PubMed
    1. Janssens JP, Howarth-frey C, Chrevolet JC, Abajo B, Rochat T. Transcutaneous PCO2 to monitor noninvasive mechanical ventilation in adults. Assessment of a new transcutaneous PCO2 device. Chest. 1998;113:768–73. - PubMed
    1. Parker SM, Gibson GJ. Evaluation of a transcutaneous carbon dioxide monitor (TOSCA) in adult patients in routine respiratory practice. Resp Med. 2007;101:261–4. - PubMed
    1. Senn O, Clarenbach CF, Kaplan V, Maggiorini M, Bloch KE. Monitoring carbon dioxide tension and arterial oxygen saturation by a single earlobe sensor in patients with critical illness or sleep apnea. Chest. 2005;128:1291–6. - PubMed
    1. Bollinger D, Steiner LA, Kasper J, Aziz OA, Filipovic M, Seeberger MD. The accuracy of non-invasive carbon dioxide monitoring: A clinical evaluation of two transcutaneous systems. Anaesthesia. 2007;62:394–9. - PubMed