Evaluation of a transcutaneous carbon dioxide monitor in patients with acute respiratory failure
- PMID: 21977067
- PMCID: PMC3183639
- DOI: 10.4103/1817-1737.84776
Evaluation of a transcutaneous carbon dioxide monitor in patients with acute respiratory failure
Abstract
Background: Non-invasive measurement of oxygenation is a routine procedure in clinical practice, but transcutaneous monitoring of PCO(2)(PtCO(2)) is used much less than expected.
Methods: The aim of our study was to analyze the value of a commercially available combined SpO(2)/PtCO(2) monitor (TOSCA-Linde Medical System, Basel, Switzerland) in adult non-invasive ventilated patients with acute respiratory failure. Eighty critically ill adult patients, requiring arterial blood sample gas analyses, underwent SpO(2) and PtCO(2) measurements (10 min after the probe was attached to an earlobe) simultaneously with arterial blood sampling. The level of agreement between PaCO(2) - PtCO(2) and SaO(2) - SpO(2)was assessed by Bland-Altman analyses.
Results: Both, SaO(2) from blood gas analysis and SpO(2) from the transcutaneous monitor, and PaCO(2) and PtCO(2) were equally useful. No measurements were outside of the acceptable clinical range of agreement of ± 7.5 mmHg.
Conclusions: The accuracy of estimation of the TOSCA transcutaneous electrode (compared with the "gold standard" blood sample gas analysis) was generally good. Moreover, TOSCA presents the advantage of the possibility of continuous non-invasive measurement. The level of agreement of the two methods of measurement allows us to state that the TOSCA sensor is useful in routine monitoring of adults admitted to an intermediate respiratory unit and undergoing non-invasive ventilation.
Keywords: Non-invasive measurement; PaCO2; PtCO2; SaO2; SpO2; respiratory failure; transcutaneous electrode.
Conflict of interest statement
Figures
References
-
- Janssens JP, Howarth-frey C, Chrevolet JC, Abajo B, Rochat T. Transcutaneous PCO2 to monitor noninvasive mechanical ventilation in adults. Assessment of a new transcutaneous PCO2 device. Chest. 1998;113:768–73. - PubMed
-
- Parker SM, Gibson GJ. Evaluation of a transcutaneous carbon dioxide monitor (TOSCA) in adult patients in routine respiratory practice. Resp Med. 2007;101:261–4. - PubMed
-
- Senn O, Clarenbach CF, Kaplan V, Maggiorini M, Bloch KE. Monitoring carbon dioxide tension and arterial oxygen saturation by a single earlobe sensor in patients with critical illness or sleep apnea. Chest. 2005;128:1291–6. - PubMed
-
- Bollinger D, Steiner LA, Kasper J, Aziz OA, Filipovic M, Seeberger MD. The accuracy of non-invasive carbon dioxide monitoring: A clinical evaluation of two transcutaneous systems. Anaesthesia. 2007;62:394–9. - PubMed