Validation of a photography-based goniometry method for measuring joint range of motion
- PMID: 21983191
- DOI: 10.1016/j.jse.2011.06.018
Validation of a photography-based goniometry method for measuring joint range of motion
Abstract
Background: A critical component of evaluating the outcomes after surgery to restore lost elbow motion is the range of motion (ROM) of the elbow. This study examined if digital photography-based goniometry is as accurate and reliable as clinical goniometry for measuring elbow ROM.
Materials and methods: Instrument validity and reliability for photography-based goniometry were evaluated for a consecutive series of 50 elbow contractures by 4 observers with different levels of elbow experience. Goniometric ROM measurements were taken with the elbows in full extension and full flexion directly in the clinic (once) and from digital photographs (twice in a blinded random manner).
Results: Instrument validity for photography-based goniometry was extremely high (intraclass correlation coefficient: extension = 0.98, flexion = 0.96). For extension and flexion measurements by the expert surgeon, systematic error was negligible (0° and 1°, respectively). Limits of agreement were 7° (95% confidence interval [CI], 5° to 9°) and -7° (95% CI, -5° to -9°) for extension and 8° (95% CI, 6° to 10°) and -7° (95% CI, -5° to -9°) for flexion. Interobserver reliability for photography-based goniometry was better than that for clinical goniometry. The least experienced observer's photographic goniometry measurements were closer to the reference measurements than the clinical goniometry measurements.
Conclusions: Photography-based goniometry is accurate and reliable for measuring elbow ROM. The photography-based method relied less on observer expertise than clinical goniometry. This validates an objective measure of patient outcome without requiring doctor-patient contact at a tertiary care center, where most contracture surgeries are done.
Copyright © 2012 Journal of Shoulder and Elbow Surgery Board of Trustees. Published by Mosby, Inc. All rights reserved.
Similar articles
-
A Comparison of Elbow Range of Motion Measurements: Smartphone-Based Digital Photography Versus Goniometric Measurements.J Hand Surg Am. 2016 Apr;41(4):510-515.e1. doi: 10.1016/j.jhsa.2016.01.006. Epub 2016 Feb 13. J Hand Surg Am. 2016. PMID: 26880499
-
Accuracy and inter-observer reliability of visual estimation compared to clinical goniometry of the elbow.Knee Surg Sports Traumatol Arthrosc. 2012 Jul;20(7):1378-85. doi: 10.1007/s00167-011-1720-9. Epub 2011 Nov 17. Knee Surg Sports Traumatol Arthrosc. 2012. PMID: 22089371
-
Are clinical photographs appropriate to determine the maximal range of motion of the knee?Acta Orthop Belg. 2010 Dec;76(6):794-8. Acta Orthop Belg. 2010. PMID: 21302578
-
The reliability and validity of goniometric elbow measurements in adults: A systematic review of the literature.Shoulder Elbow. 2018 Oct;10(4):274-284. doi: 10.1177/1758573218774326. Epub 2018 Jun 3. Shoulder Elbow. 2018. PMID: 30214494 Free PMC article. Review.
-
Smartphone Applications for Range of Motion Measurement in Clinical Practice: A Systematic Review.Stud Health Technol Inform. 2020 Jun 16;270:1389-1390. doi: 10.3233/SHTI200456. Stud Health Technol Inform. 2020. PMID: 32570673
Cited by
-
Posteriorly Displaced Radial Head Fractures May Represent the Footprint of an Elbow Dislocation or Subluxation as a Variant of Modified Mason Type 4.Arch Bone Jt Surg. 2022 Jun;10(6):501-506. doi: 10.22038/ABJS.2021.55486.2764. Arch Bone Jt Surg. 2022. PMID: 35928906 Free PMC article.
-
Smartphone applications for the evaluation of pathologic shoulder range of motion and shoulder scores-a comparative study.JSES Open Access. 2018 Mar 13;2(1):109-114. doi: 10.1016/j.jses.2017.10.001. eCollection 2018 Mar. JSES Open Access. 2018. PMID: 30675577 Free PMC article.
-
Extrinsic feedback from a feedback device promotes the learning of range of motion measurements.J Phys Ther Sci. 2020 Feb;32(2):114-119. doi: 10.1589/jpts.32.114. Epub 2020 Feb 14. J Phys Ther Sci. 2020. PMID: 32158073 Free PMC article.
-
The Virtual Shoulder Physical Exam.HSS J. 2021 Feb;17(1):59-64. doi: 10.1177/1556331620975033. Epub 2021 Feb 21. HSS J. 2021. PMID: 33967643 Free PMC article. No abstract available.
-
Validity Analysis of Monocular Human Pose Estimation Models Interfaced with a Mobile Application for Assessing Upper Limb Range of Motion.Sensors (Basel). 2024 Dec 14;24(24):7983. doi: 10.3390/s24247983. Sensors (Basel). 2024. PMID: 39771719 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources