A review of the analytic performance of the Reflotron System for cholesterol testing
- PMID: 2199060
A review of the analytic performance of the Reflotron System for cholesterol testing
Abstract
The analytic performance of the Reflotron System for cholesterol testing has been assessed in 17 studies since 1987. The precision of the test (total coefficient of variation) ranged from 0.8% to 9.9% and was between 1% and 5% in 23 of the 27 specimen types tested. The correlation coefficient of the Reflotron with that of a reference methods was over .9 in 15 specimen types tested and over .95 in ten of 15. The type of specimen tested and the training of the test personnel affected both precision and accuracy. It is concluded that the Reflotron analyzer provides reliable and accurate cholesterol measurements.
Similar articles
-
Evaluation of within-day precision of serum cholesterol measured by a portable analyzer.Med Sci Sports Exerc. 1992 Jan;24(1):134-8. Med Sci Sports Exerc. 1992. PMID: 1548987
-
[Measurement of cholesterol in capillary blood using the Reflotron system. Results from approximately 1,000 comparisons with reference measurements of cholesterol in venous serum].Ugeskr Laeger. 1990 Jun 11;152(24):1739-43. Ugeskr Laeger. 1990. PMID: 2360292 Danish.
-
How accurately does the Reflotron dry-chemistry system measure plasma total cholesterol levels when used a community-screening device.Med J Aust. 1988 Aug 1;149(3):122-5. Med J Aust. 1988. PMID: 3398798
-
The Centers for Disease Control-National Heart, Lung and Blood Institute Lipid Standardization Program. An approach to accurate and precise lipid measurements.Clin Lab Med. 1989 Mar;9(1):105-35. Clin Lab Med. 1989. PMID: 2538292 Review.
-
Laboratory testing and biochemical analysis of hyperlipidaemias.Postgrad Med J. 1993;69 Suppl 1:S12-7. Postgrad Med J. 1993. PMID: 8497452 Review.
Cited by
-
Development and Validation of a Personalized, Sex-Specific Prediction Algorithm of Severe Atheromatosis in Middle-Aged Asymptomatic Individuals: The ILERVAS Study.Front Cardiovasc Med. 2022 Jul 14;9:895917. doi: 10.3389/fcvm.2022.895917. eCollection 2022. Front Cardiovasc Med. 2022. PMID: 35928938 Free PMC article.
-
Comparison of the use of a dry chemistry analyser in primary care in Norway and the United Kingdom.Br J Gen Pract. 1993 Jan;43(366):10-4. Br J Gen Pract. 1993. PMID: 8457356 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Medical