Application of the gamma evaluation method in Gamma Knife film dosimetry
- PMID: 21992392
- DOI: 10.1118/1.3641644
Application of the gamma evaluation method in Gamma Knife film dosimetry
Abstract
Purpose: Gamma Knife (GK) radiosurgery is a minimally invasive surgical technique for the treatment of intracranial lesions. To minimize neurological deficits, submillimeter accuracy is required during treatment delivery. In this paper, the delivery accuracy of GK radiosurgery was assessed with the gamma evaluation method using planning dose distribution and film measurement data.
Methods: Single 4, 8, and 16 mm and composite shot plans were developed for evaluation using the GK Perfexion (PFX) treatment planning system (TPS). The planning dose distributions were exported as digital image communications in medicine - radiation therapy (DICOM RT) files using a new function of GK TPS. A maximum dose of 8 Gy was prescribed for four test plans. Irradiation was performed onto a spherical solid water phantom using Gafchromic EBT2 films in the axial and coronal planes. The exposed films were converted to absolute dose based on a 4th-order polynomial calibration curve determined using ten calibration films. The film measurement results and planning dose distributions were registered for further analysis in the same Leksell coordinate using in-house software. The gamma evaluation method was applied to two dose distributions with varying spatial tolerance (0.3-2.0 mm) and dosimetric tolerance (0.3-2.0%), to verify the accuracy of GK radiosurgery. The result of gamma evaluation was assessed using pass rate, dose gamma index histogram (DGH), and dose pass rate histogram (DPH).
Results: The 20, 50, and 80% isodose lines found in film measurements were in close agreement with the planning isodose lines, for all dose levels. The comparison of diagonal line profiles across the axial plane yielded similar results. The gamma evaluation method resulted in high pass rates of >95% within the 50% isodose line for 0.5 mm∕0.5% tolerance criteria, in both the axial and coronal planes. They satisfied 1.0 mm∕1.0% criteria within the 20% isodose line. Our DGH and DPH also showed that low isodose lines exhibited inferior gamma indexes and pass rates compared with higher isodose lines.
Conclusions: The gamma evaluation method was applicable to GK radiosurgery. For all test plans, planning dose distribution and film measurement met the tolerance criteria of 0.5 mm∕0.5% within the 50% isodose line which are used for marginal dose prescription.
Similar articles
-
Verification of dose profiles generated by the convolution algorithm of the gamma knife® radiosurgery planning system.Med Phys. 2017 Sep;44(9):4880-4889. doi: 10.1002/mp.12347. Epub 2017 Jul 28. Med Phys. 2017. PMID: 28513854
-
[Validation and analysis of dose distributions in a new and entirely redesigned cobalt-60 stereotactic radiosurgery units].Nihon Hoshasen Gijutsu Gakkai Zasshi. 2015 Feb;71(2):92-8. doi: 10.6009/jjrt.2015_JSRT_71.2.92. Nihon Hoshasen Gijutsu Gakkai Zasshi. 2015. PMID: 25748009 Japanese.
-
Verification of Gamma Knife extend system based fractionated treatment planning using EBT2 film.Med Phys. 2013 Dec;40(12):122104. doi: 10.1118/1.4832138. Med Phys. 2013. PMID: 24320531
-
3-D treatment planning system-Leksell Gamma Knife treatment planning system.Med Dosim. 2018 Summer;43(2):177-183. doi: 10.1016/j.meddos.2018.03.001. Epub 2018 Apr 12. Med Dosim. 2018. PMID: 29656889 Review.
-
[Investigating the clinical applicability of EBT2 self-developing films].Magy Onkol. 2016 Nov 29;60(4):299-304. Epub 2016 May 25. Magy Onkol. 2016. PMID: 27898748 Review. Hungarian.
Cited by
-
Theoretical study of doped porous silicon in cantor quasi periodic structure for gamma radiation detection.Sci Rep. 2025 Apr 29;15(1):14995. doi: 10.1038/s41598-025-94555-4. Sci Rep. 2025. PMID: 40301423 Free PMC article.
-
Assessment of the accuracy and stability of frameless gamma knife radiosurgery.J Appl Clin Med Phys. 2018 Jul;19(4):148-154. doi: 10.1002/acm2.12365. Epub 2018 Jun 3. J Appl Clin Med Phys. 2018. PMID: 29862671 Free PMC article.
-
MAGAT gel and EBT2 film-based dosimetry for evaluating source plugging-based treatment plan in Gamma Knife stereotactic radiosurgery.J Appl Clin Med Phys. 2012 Nov 8;13(6):3877. doi: 10.1120/jacmp.v13i6.3877. J Appl Clin Med Phys. 2012. PMID: 23149780 Free PMC article.
-
Photonic crystal with a defect layer of silicon containing polymer nanocomposites as radiation detector.Sci Rep. 2025 Mar 7;15(1):7935. doi: 10.1038/s41598-025-91050-8. Sci Rep. 2025. PMID: 40050332 Free PMC article.
-
Tuning of AcurosXB source size setting for small intracranial targets.J Appl Clin Med Phys. 2017 May;18(3):170-181. doi: 10.1002/acm2.12091. Epub 2017 May 4. J Appl Clin Med Phys. 2017. PMID: 28470819 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Miscellaneous