Ultrasound imaging of the thoracic spine in paramedian sagittal oblique plane: the correlation between estimated and actual depth to the epidural space
- PMID: 22005658
- DOI: 10.1097/AAP.0b013e31823217e7
Ultrasound imaging of the thoracic spine in paramedian sagittal oblique plane: the correlation between estimated and actual depth to the epidural space
Abstract
Background: Ultrasound (US) imaging of the spine has been shown to be a reliable tool to facilitate lumbar epidural needle placement; however, its feasibility in thoracic epidural placement is still unknown. The objective of this study was to assess the accuracy and reliability of prepuncture US imaging in the paramedian sagittal oblique plane to estimate the depth to the epidural space and optimum insertion point for guiding epidural needle placement at the mid-low thoracic level.
Methods: This prospective study included 35 healthy adult patients who requested thoracic epidural analgesia before their upper abdominal surgeries. Ultrasound imaging was done in the paramedian sagittal oblique plane at the desired thoracic level to identify the intervertebral space, the distance from the skin to the epidural space (US depth [UD]) and the needle insertion point. Subsequently, a staff anesthesiologist located the epidural space through the predetermined insertion point and marked the actual distance from the skin to the epidural space (needle depth [ND]) on the needle with a sterile marker. The agreement between the UD and the ND was calculated using the Pearson and concordance correlation coefficients and Bland-Altman analysis with 95% limits of agreement.
Results: The average patient age was 56 (SD, 14) years, and body mass index was 28 (SD, 6) kg/m2. The precision of the agreement between UD and ND estimated by Pearson correlation coefficient was 0.75, and the accuracy was 0.80, whereas the concordance correlation coefficient was 0.60 (confidence interval, 0.43-0.78). The mean UD and ND were 4.3 (SD, 0.96) and 5.0 (SD, 1.2) cm, respectively. The Bland-Altman analysis showed a mean difference of -0.71 cm (95% limits of agreement, 0.8 to -2.2 cm). There was a significant direct correlation of the ND with the body mass index (r2 = 0.27, P = 0.008). The mean number of attempts was 1 (p25-p75 = 1-2), and the epidural space was identified with 2 or less redirections in 88% of the cases.
Conclusions: We found a good correlation between the US-estimated distance to the epidural space and the actual measured needle distance in our patients. We suggest that our proposed prepuncture US method, using the paramedian sagittal oblique approach, can be a useful guide to facilitate the placement of epidural needles at mid-low thoracic levels. A randomized controlled trial is necessary to confirm the utility of prepuncture US in thoracic epidural placement.
Similar articles
-
Ultrasound imaging of the lumbar spine in the transverse plane: the correlation between estimated and actual depth to the epidural space in obese parturients.Anesth Analg. 2009 Jun;108(6):1876-81. doi: 10.1213/ane.0b013e3181a323f6. Anesth Analg. 2009. PMID: 19448216
-
Ultrasound estimates for midline epidural punctures in the obese parturient: paramedian sagittal oblique is comparable to transverse median plane.Anesth Analg. 2013 Apr;116(4):829-35. doi: 10.1213/ANE.0b013e31827f55f0. Epub 2013 Feb 5. Anesth Analg. 2013. PMID: 23385058
-
Ultrasound using the transverse approach to the lumbar spine provides reliable landmarks for labor epidurals.Anesth Analg. 2007 May;104(5):1188-92, tables of contents. doi: 10.1213/01.ane.0000250912.66057.41. Anesth Analg. 2007. PMID: 17456672
-
Caudal and epidural blocks in infants and small children: historical perspective and ultrasound-guided approaches.Korean J Anesthesiol. 2018 Dec;71(6):430-439. doi: 10.4097/kja.d.18.00109. Epub 2018 Aug 8. Korean J Anesthesiol. 2018. PMID: 30086609 Free PMC article. Review.
-
Conventional anatomical landmark versus preprocedural ultrasound for thoracic epidural analgesia: A systematic review and meta-analysis.J Perioper Pract. 2024 Oct;34(10):315-325. doi: 10.1177/17504589231181974. Epub 2023 Sep 13. J Perioper Pract. 2024. PMID: 37702201
Cited by
-
Real-time ultrasound-guided versus anatomic landmark-based thoracic epidural placement: a prospective, randomized, superiority trial.BMC Anesthesiol. 2022 Jun 25;22(1):198. doi: 10.1186/s12871-022-01730-5. BMC Anesthesiol. 2022. PMID: 35752755 Free PMC article. Clinical Trial.
-
Effectiveness of preoperative thoracic epidural testing strategies: a retrospective comparison of three commonly used testing methods.Can J Anaesth. 2024 Jun;71(6):793-801. doi: 10.1007/s12630-023-02545-9. Epub 2023 Jul 28. Can J Anaesth. 2024. PMID: 37505418 English.
-
3D Ultrasound Imaging of the Spine.Adv Exp Med Biol. 2022;1364:349-372. doi: 10.1007/978-3-030-91979-5_16. Adv Exp Med Biol. 2022. PMID: 35508883
-
A randomized comparison of loss of resistance versus loss of resistance plus electrical stimulation: effect on success of thoracic epidural placement.BMC Anesthesiol. 2022 Feb 9;22(1):43. doi: 10.1186/s12871-022-01584-x. BMC Anesthesiol. 2022. PMID: 35139802 Free PMC article. Clinical Trial.
-
Abdominal girth has a strong correlation with actual and ultrasound estimated epidural depth.Turk J Med Sci. 2019 Dec 16;49(6):1715-1720. doi: 10.3906/sag-1902-115. Turk J Med Sci. 2019. PMID: 31655518 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Research Materials