The effect of pulling out cochlear implant electrodes on inner ear microstructures: a temporal bone study
- PMID: 22007225
- PMCID: PMC3191782
- DOI: 10.1155/2011/107176
The effect of pulling out cochlear implant electrodes on inner ear microstructures: a temporal bone study
Abstract
The exchange of an cochlear implant or the re-positioning of an electrode have become more frequently required than a decade ago. The consequences of such procedures at a microstructural level within the cochlea are not known. It was the aim of the present study to further investigate the effects of an CI electrode pull-out. Therefore 10 freshly harvested temporal bones (TB) were histologically evaluated after a cochlear implant electrode pull-out of a perimodiolar electrode. In additional 9 TB the intrascalar movements of the CI electrode while being pulled-out were digitally analysed by video- capturing. Histologically, a disruption of the modiolar wall or the spiral osseous lamina were not observed. In one TB, a basilar membrane lifting up was found, but it could not be undoubtedly attributed to the pull-out of the electrode. When analyzing the temporal sequence of the electrode movement during the pull-out, the electrode turned in one case so that the tip elevates the basilar membrane. The pull- out of perimodiolarly placed CI electrodes does not damage the modiolar wall at a microstructural level and should be guided (e.g., forceps) to prevent a 90 o turning of the electrode tip into the direction of the basilar membrane.
Figures
References
-
- Eshraghi AA, Yang NW, Balkany TJ. Comparative study of cochlear damage with three perimodiolar electrode designs. Laryngoscope. 2003;113(3):415–419. - PubMed
-
- Wardrop P, Whinney D, Rebscher SJ, Roland JT, Jr., Luxford W, Leake PA. A temporal bone study of insertion trauma and intracochlear position of cochlear implant electrodes. I: comparison of Nucleus banded and Nucleus Contour™ electrodes. Hearing Research. 2005;203(1-2):54–67. - PubMed
-
- Wardrop P, Whinney D, Rebscher SJ, Luxford W, Leake P. A temporal bone study of insertion trauma and intracochlear position of cochlear implant electrodes. II: comparison of Spiral Clarion™ and HiFocus II™ electrodes. Hearing Research. 2005;203(1-2):68–79. - PubMed
-
- Klenzner T, Richter B, Nagursky H, Schipper J, Laszig R, Aschendorff A. Evaluation of the inserfion-trauma of the Nucleus® Contour Advance™ electrode-array in a human temporal bone model. Laryngo-Rhino-Otologie. 2004;83(12):840–844. - PubMed
-
- Wright CG, Roland PS, Kuzma J. Advanced bionics thin lateral and helix II electrodes: a temporal bone study. Laryngoscope. 2005;115(11):2041–2045. - PubMed
LinkOut - more resources
Full Text Sources