. 2012 Apr;22(4):765-71.
doi: 10.1007/s00330-011-2308-y. Epub 2011 Oct 20.

Three-dimensional measurements of the lower extremity in children and adolescents using a low-dose biplanar X-ray device

Affiliations

Three-dimensional measurements of the lower extremity in children and adolescents using a low-dose biplanar X-ray device

Ramon Gheno et al. Eur Radiol. 2012 Apr.

Abstract

Objective: To evaluate three-dimensional (3D) measurements of the lower extremity using a biplanar low-dose X-ray device in children and adolescents.

Methods: Firstly, 3D measurements of eight dried bones were analysed by a biplanar low-dose X-ray device (LDX) using stereoscopic software and compared with 3D computed tomography (CT). Secondly, 47 lower limbs of children and adolescents were studied using LDX two-dimensional (2D) and 3D measurements. Both parts were evaluated for femoral and tibial lengths and mechanical angles, frontal and lateral knee angulations, and the femoral neck-shaft angle.

Results: The 3D specimen comparison between LDX and CT measurements showed no significant differences: femoral length (P = 0.069), tibial length (P = 0.059), femoral mechanical angle (P = 0.475), tibial mechanical angle (P = 0.067), frontal knee angulation (P = 0.198), lateral knee angulation (P = 0.646) and femoral neck-shaft angle (P = 0.068). The comparison between LDX 2D and 3D measurements showed significant differences in tibial length (P = 0.003), femoral mechanical angle (P < 0.001) and femoral neck-shaft angle (P = 0.001); other parameters were unremarkable.

Conclusions: The 3D LDX system presented reliable measurements compared with 3D CT. Differences between LDX 2D and 3D measurements were noted in the femoral mechanical angle, femoral neck-shaft angle and tibial length. Moderate to good interobserver agreement for the 3D LDX measurements were found.

Key points: Low radiation dose is essential when assessing potential lower extremity discrepancies. A new biplanar low-dose X-ray device can assess such discrepancies in children/adolescents. This LDX device provides equally reliable 3D measurements as prevalent practice LDX measurements carry good overall interobserver agreement.

PubMed Disclaimer

References

    1. J Rheumatol. 2007 Sep;34(9):1796-801 - PubMed
    1. AJR Am J Roentgenol. 2008 Feb;190(2):273-4 - PubMed
    1. Can J Surg. 2006 Dec;49(6):401-6 - PubMed
    1. Gait Posture. 2002 Apr;15(2):195-206 - PubMed
    1. J Bone Joint Surg Am. 1997 Nov;79(11):1690-8 - PubMed

Publication types

LinkOut - more resources