Comparative evaluation of the QIAGEN QIAsymphony® SP system and bioMérieux NucliSens easyMAG automated extraction platforms in a clinical virology laboratory
- PMID: 22014618
- DOI: 10.1016/j.jcv.2011.08.016
Comparative evaluation of the QIAGEN QIAsymphony® SP system and bioMérieux NucliSens easyMAG automated extraction platforms in a clinical virology laboratory
Abstract
Background: The QIAGEN QIAsymphony(®) SP is an automated system that can process a variety of different sample types for nucleic acid extraction.
Objectives: To compare this system against the bioMérieux NucliSens easyMAG using a range of common clinical sample types.
Study design: Nucleic acid extracts were tested on 6 in-house real time viral PCR assays: quantitative cytomegalovirus (CMV); respiratory and enteric multiplex (10 viruses); influenza A H1N1 2009 specific H1 and N1 assays (sH1/N1); herpes simplex virus (HSV) 1, HSV 2 and varicella zoster virus (VZV) multiplex; norovirus genogroups I and II (Noro GI/II) multiplex.
Results: Extraction of the clinical sample by either QIAsymphony(®) SP or NucliSens easyMAG gave similar results for each PCR; CMV viral loads, 52 plasma samples had a mean difference (easyMAG-QIAsymphony(®)) of 0.002 log(10)copies/ml (s.d. 0.536), 52 whole blood samples had a mean difference of -0.232 log(10)copies/ml (s.d. 0.490). Concordance for the qualitative assays were; 64/67 (95.5%) for the respiratory and enteric multiplex, all 28 (100%) for the sH1, sN1 and influenza A matrix multiplex, 33/34 (97%) for the HSV1/HSV2/VZV multiplex and all 15 (100%) for the Noro GI/II. Inter- and intra-run variation, determined for a 10-fold dilution series of CMV (5.20-3.20 log(10)copies/ml), was less than 0.63 log(10)copies/ml.
Conclusions: Our evaluation found the performance of the QIAsymphony(®) SP comparable to the NucliSens easyMAG for a range of sample types commonly extracted in a clinical virology laboratory. In total, 331/343 (96.5%) PCR results were concordant on samples extracted by both platforms.
Copyright © 2011 Elsevier B.V. All rights reserved.
Similar articles
-
Extraction of viral nucleic acids: comparison of five automated nucleic acid extraction platforms.J Clin Virol. 2012 Jul;54(3):255-9. doi: 10.1016/j.jcv.2012.03.008. Epub 2012 Apr 13. J Clin Virol. 2012. PMID: 22503856
-
Comparison of the QIAsymphony automated nucleic acid extraction and PCR setup platforms with NucliSens easyMAG and Corbett CAS-1200 liquid handling station for the detection of enteric pathogens in fecal samples.J Microbiol Methods. 2011 Feb;84(2):335-40. doi: 10.1016/j.mimet.2010.12.024. Epub 2010 Dec 31. J Microbiol Methods. 2011. PMID: 21199666
-
Performance evaluation of the automated NucliSens easyMAG nucleic acid extraction platform in comparison with QIAamp Mini kit from clinical specimens.Diagn Microbiol Infect Dis. 2009 Jun;64(2):158-65. doi: 10.1016/j.diagmicrobio.2009.02.013. Diagn Microbiol Infect Dis. 2009. PMID: 19500527
-
Molecular diagnosis of respiratory viruses.Future Microbiol. 2010 Jun;5(6):901-16. doi: 10.2217/fmb.10.48. Future Microbiol. 2010. PMID: 20521935 Review.
-
Next-generation sequencing technologies in diagnostic virology.J Clin Virol. 2013 Oct;58(2):346-50. doi: 10.1016/j.jcv.2013.03.003. Epub 2013 Mar 21. J Clin Virol. 2013. PMID: 23523339 Review.
Cited by
-
Comparative evaluation of the QIAsymphony RGQ system with the easyMAG/R-gene combination for the quantitation of cytomegalovirus DNA load in whole blood.Virol J. 2012 Oct 9;9:231. doi: 10.1186/1743-422X-9-231. Virol J. 2012. PMID: 23046712 Free PMC article.
-
Laboratory detection of respiratory viruses by automated techniques.Open Virol J. 2012;6:151-9. doi: 10.2174/1874357901206010151. Epub 2012 Nov 30. Open Virol J. 2012. PMID: 23248735 Free PMC article.
-
Performance of the MagNA Pure 96 system for cytomegalovirus nucleic acid amplification testing in clinical samples.J Clin Microbiol. 2013 May;51(5):1600-1. doi: 10.1128/JCM.03289-12. Epub 2013 Mar 6. J Clin Microbiol. 2013. PMID: 23467597 Free PMC article.
-
Automated Nucleic Acid Extraction Systems for Detecting Cytomegalovirus and Epstein-Barr Virus Using Real-Time PCR: A Comparison Study Between the QIAsymphony RGQ and QIAcube Systems.Ann Lab Med. 2017 Mar;37(2):129-136. doi: 10.3343/alm.2017.37.2.129. Ann Lab Med. 2017. PMID: 28028999 Free PMC article.
-
Fully automated quantification of cytomegalovirus (CMV) in whole blood with the new sensitive Abbott RealTime CMV assay in the era of the CMV international standard.J Clin Microbiol. 2013 Jul;51(7):2096-102. doi: 10.1128/JCM.00067-13. Epub 2013 Apr 24. J Clin Microbiol. 2013. PMID: 23616450 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical