Prolapsed double-canted bipolar left ventricular lead for pacing the left atrium via the coronary sinus: experience in 11 patients
- PMID: 22016306
- DOI: 10.1093/europace/eur331
Prolapsed double-canted bipolar left ventricular lead for pacing the left atrium via the coronary sinus: experience in 11 patients
Abstract
Aims: High thresholds and frequent lead dislodgement limit pacing the left atrium (LA) from the mid to distal coronary sinus (CS). The aim of this report is to describe a method for and the results of prolapsing a double-canted bipolar lead into the mid-to-distal CS to eliminate lead dislodgement and improve pacing thresholds.
Methods and results: After CS access the 9 Fr. anatomic sheath is withdrawn to the right atrium (RA) over an extra support wire. A double-canted bipolar lead is advanced into the RA until the proximal bend is outside the tip of the sheath. With the stylet withdrawn to the proximal bend, the sheath and lead are advanced over the wire back into the CS. The lead distal to the proximal bend is prolapsed beside the sheath as the tip of the sheath enters the CS. The lead was successfully prolapsed in 11 consecutive patients. In one patient, capture was >5 V in all locations. Of the 10 successful implants, the acute thresholds were: mean 1.53 V, median 1.35 V, range 0.4-4.0 V. Chronic thresholds were: mean 2 V, median 2 V range 0.4-4.0 V. There were no displaced leads or lead fractures through 6-10 months of follow-up.
Conclusions: Prolapse of a commercially available double-canted bipolar passive fixation lead eliminates lead dislodgment and improves thresholds providing a means for permanent pacing of the LA from the mid to distal CS and provides the design principles for a dedicated lead.
Similar articles
-
A comparison between passive and active fixation leads in the coronary sinus for biatrial pacing: initial experience.Europace. 2000 Jul;2(3):228-32. doi: 10.1053/eupc.2000.0105. Europace. 2000. PMID: 11227593 Clinical Trial.
-
Permanent left atrial pacing with a specifically designed coronary sinus lead.Pacing Clin Electrophysiol. 1997 Nov;20(11):2755-64. doi: 10.1111/j.1540-8159.1997.tb05433.x. Pacing Clin Electrophysiol. 1997. PMID: 9392806
-
Surgical epicardial left ventricular lead versus coronary sinus lead placement in biventricular pacing.Eur J Cardiothorac Surg. 2005 Feb;27(2):235-42. doi: 10.1016/j.ejcts.2004.09.029. Eur J Cardiothorac Surg. 2005. PMID: 15691676
-
Multi-Axes Lead With Tetrahedral Electrode Tip for Cardiac-Implantable Devices: Creative Concept for Pacing and Sensing Technology.Can J Cardiol. 2021 Nov;37(11):1808-1817. doi: 10.1016/j.cjca.2021.07.013. Epub 2021 Jul 29. Can J Cardiol. 2021. PMID: 34333028 Review.
-
Challenging Implants Require Tools and Techniques Not Tips and Tricks.Card Electrophysiol Clin. 2019 Mar;11(1):75-87. doi: 10.1016/j.ccep.2018.11.003. Card Electrophysiol Clin. 2019. PMID: 30717855 Review.
Cited by
-
Interatrial conduction block-related atrioventricular dyssynchrony treated with dual-site atrial pacing.HeartRhythm Case Rep. 2021 Jan 21;7(4):224-228. doi: 10.1016/j.hrcr.2021.01.004. eCollection 2021 Apr. HeartRhythm Case Rep. 2021. PMID: 34026501 Free PMC article. No abstract available.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources