Nonphysiologic noise early after defibrillator implantation in Canada: incidence and implications: a report from the Canadian Heart Rhythm Society Device Committee

Affiliations
Case Reports

Nonphysiologic noise early after defibrillator implantation in Canada: incidence and implications: a report from the Canadian Heart Rhythm Society Device Committee

Vidal Essebag et al. Heart Rhythm. 2012 Mar.

Abstract

Background: Following recent cases of nonphysiologic noise noted early after defibrillator implantation, the Canadian Heart Rhythm Society Device Committee decided to evaluate the implications of this problem.

Objective: To determine the incidence and consequences of nonphysiologic noise early after defibrillator implantation.

Methods: The Canadian Heart Rhythm Society Device Committee surveyed all Canadian defibrillator implanting centers regarding their implant volumes and number of cases where nonphysiologic noise had been noted early (< 24 hours) after implant over the preceding 2 years. For such cases, information regarding the manufacturer and occurrence of inappropriate shock or inhibition of pacing was reported.

Results: Responses were obtained from 20 of 23 surveyed implanting centers, with a total implant volume of 4960 defibrillators per year. The occurrence of nonphysiologic noise early after implantation was noted in 25 cases over the preceding 2 years (0.25%). Noise was detected in devices of all 3 of the leading volume device manufacturers. There were 2 cases of inappropriate shocks and 2 cases of symptomatic pacing inhibition. In 4 cases, removal of the lead from the header and retesting with the analyzer confirmed normal lead function. In all cases, the noise resolved within 24 hours.

Conclusion: Nonphysiologic noise early after defibrillator implantation was noted in 0.25% of procedures and was not limited to one specific manufacturer. This noise may result in an inappropriate shock or inhibition of pacing in a pacemaker-dependent patient. This transient phenomenon, possibly related to fluid and/or air in the header, appears to always resolve without surgical intervention.