Iterative reconstruction in head CT: image quality of routine and low-dose protocols in comparison with standard filtered back-projection
- PMID: 22033719
- PMCID: PMC7964820
- DOI: 10.3174/ajnr.A2749
Iterative reconstruction in head CT: image quality of routine and low-dose protocols in comparison with standard filtered back-projection
Abstract
Background and purpose: IR has recently demonstrated its capacity to reduce noise and permit dose reduction in abdominal and thoracic CT applications. The purpose of our study was to assess the potential benefit of IR in head CT by comparing objective and subjective image quality with standard FBP at various dose levels.
Materials and methods: Ninety consecutive patients were randomly assigned to undergo nonenhanced and contrast-enhanced head CT at a standard dose (320 mAs; CTDI, 60.1) or 15% (275 mAs; CTDI, 51.8) and 30% (225 mAs; CTDI, 42.3) dose reduction. All acquisitions were reconstructed with IR in image space, and FBP and images were assessed in terms of quantitative and qualitative IQ.
Results: Compared with FBP, IR resulted in lower image noise (P ≤ .02), higher CNR (P ≤ .03), and improved subjective image quality (P ≤ .002) at all dose levels. While degradation of objective and subjective IQ at 15% dose reduction was fully compensated by IR (CNR, 1.98 ± 0.4 at 320 mAs with FBP versus 2.05 ± 0.4 at 275 mAs with IR; IQ, 1.8 versus 1.7), IQ was considerably poorer at 70% standard dose despite using the iterative approach (CNR, 1.98 ± 0.3 at 320 mAs with FBP versus 1.85 ± 0.4 at 225 mAs with IR, P = .18; IQ, 1.8 versus 2.2, P = .03). Linear regression analysis of CNR against tube current suggests that standard CNR may be obtained until approximately 20.4% dose reduction when IR is used.
Conclusions: Compared with conventional FBP, IR of head CT is associated with significant improvement of objective and subjective IQ and may allow dose reductions in the range of 20% without compromising standard image quality.
Figures
References
-
- Brenner DJ, Hall EJ. Computed tomography: an increasing source of radiation exposure. N Engl J Med 2007; 357: 2277– 84 - PubMed
-
- Brenner DJ. Slowing the increase in the population dose resulting from CT scans. Radiat Res 2010; 174: 809– 15 - PubMed
-
- Mahesh M. NCRP report number 160: its significance to medical imaging. J Am Coll Radiol 2009; 6: 890– 92 - PubMed
-
- Kalra MK, Maher MM, Blake MA, et al. Detection and characterization of lesions on low-radiation-dose abdominal CT images postprocessed with noise reduction filters. Radiology 2004; 232: 791– 97 - PubMed
-
- Kalra MK, Maher MM, Toth TL, et al. Techniques and applications of automatic tube current modulation for CT. Radiology 2004; 233: 649– 57 - PubMed
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Miscellaneous