Using the inflating syringe as a safety valve to limit laryngeal mask airway cuff pressure
- PMID: 22076616
- DOI: 10.1007/s10877-011-9319-8
Using the inflating syringe as a safety valve to limit laryngeal mask airway cuff pressure
Abstract
Objective: Hyperinflation of the laryngeal mask airway (LMA) cuff is thought to be the etiology underlying many of the complications associated with the use of this device. Until now, there has not been a clinically acceptable method (besides direct measurement) to assure that the cuff pressure is maintained less than the recommended maximum value of 44 mm Hg (60 cm H(2)O).
Methods: We inflated sizes #2 and #5 LMAs with air to 40, 60, or 120 mm Hg starting pressures, using 30- and 60-ml BD™ and B Braun™ syringes; we then allowed the syringe plungers to recoil to equilibrium before removing the syringe from the LMA inflation port. Residual LMA cuff pressures following complete passive recoil were measured and recorded.
Results: A number of combinations of syringes (30 and 60 ml) and starting pressures (40, 60, 120 mm Hg) resulted in safe residual (#2 and #5 LMA) cuff pressures of <44 mm Hg.
Conclusion: When using specific combinations of syringes, LMA sizes and inflation pressures, these data demonstrate an efficient, practical and easy method to achieve an initial equilibrium recoil LMA cuff pressure that is less than, or very near to, the recommended upper safe limit of 44 mm Hg.
Comment in
-
Using a modified syringe technique to adjust the intracuff pressure of a laryngeal mask airway.Acta Anaesthesiol Taiwan. 2015 Dec;53(4):146-7. doi: 10.1016/j.aat.2015.03.004. Epub 2015 May 8. Acta Anaesthesiol Taiwan. 2015. PMID: 25962713
Similar articles
-
Clinical application of limiting laryngeal mask airway cuff pressures utilizing inflating syringe intrinsic recoil.Rom J Anaesth Intensive Care. 2018 Apr;25(1):11-18. doi: 10.21454/rjaic.7518.251.cuf. Rom J Anaesth Intensive Care. 2018. PMID: 29756057 Free PMC article.
-
Laryngeal mask airways--to inflate or to deflate after insertion?Paediatr Anaesth. 2009 Sep;19(9):837-43. doi: 10.1111/j.1460-9592.2009.03109.x. Paediatr Anaesth. 2009. PMID: 19691691
-
Correlation and variation of cuff inflating volumes and pressures in different adult models of laryngeal mask: a prospective randomized trial.BMC Anesthesiol. 2020 May 7;20(1):108. doi: 10.1186/s12871-020-01028-4. BMC Anesthesiol. 2020. PMID: 32380954 Free PMC article. Clinical Trial.
-
Clinical evaluation of a novel LMA with a color-coded pressure gauge.Int J Pediatr Otorhinolaryngol. 2013 Jan;77(1):76-8. doi: 10.1016/j.ijporl.2012.09.036. Epub 2012 Oct 23. Int J Pediatr Otorhinolaryngol. 2013. PMID: 23098746
-
Cuff filling volumes and pressures in pediatric laryngeal mask airways.Paediatr Anaesth. 2006 Jan;16(1):25-30. doi: 10.1111/j.1460-9592.2005.01672.x. Paediatr Anaesth. 2006. PMID: 16409525
Cited by
-
The effect of esophagogastroduodenoscopy probe insertion on the intracuff pressure of airway devices in children during general anesthesia.J Anesth. 2017 Apr;31(2):278-281. doi: 10.1007/s00540-016-2297-8. Epub 2016 Dec 21. J Anesth. 2017. PMID: 28004199
-
Clinical application of limiting laryngeal mask airway cuff pressures utilizing inflating syringe intrinsic recoil.Rom J Anaesth Intensive Care. 2018 Apr;25(1):11-18. doi: 10.21454/rjaic.7518.251.cuf. Rom J Anaesth Intensive Care. 2018. PMID: 29756057 Free PMC article.
-
Spring recoil and supraglottic airway devices: lessons from the law of conservation of energy.Rom J Anaesth Intensive Care. 2018 Apr;25(1):7-9. doi: 10.21454/rjaic.7518.251.sor. Rom J Anaesth Intensive Care. 2018. PMID: 29756056 Free PMC article. No abstract available.
References
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical