Keratometry evaluations with the Pentacam high resolution in comparison with the automated keratometry and conventional corneal topography
- PMID: 22081146
- DOI: 10.1097/ICO.0b013e318204c666
Keratometry evaluations with the Pentacam high resolution in comparison with the automated keratometry and conventional corneal topography
Abstract
Purpose: To determine the reliability and repeatability of keratometry (K) measurements obtained with the Pentacam high resolution (HR), automated keratometry, and corneal topography systems.
Methods: The right eyes of 46 healthy subjects were examined prospectively. Keratometry measurements in the flat (Kf) and steep (Ks) meridians were taken by 2 independent investigators with the Pentacam HR (Oculus, Wetzlar, Germany) followed by automated keratorefractometry (KR-8100; Topcon, Tokyo, Japan), and corneal topography (TMS-4; Tomey, Erlangen, Germany).
Results: The mean K readings of the Pentacam HR, automated keratometry, and corneal topography were 43.40/43.34 diopter (D), 43.99/43.98 D, and 43.80/43.83 D, respectively. The difference between the values was statistically significant (P < 0.0001, repeated measures analysis of variance). Strong significant correlation was observed between the Pentacam HR and keratometry (Kf: r = 0.952/0.954; Ks: r = 0.845, Spearman rank test), and Pentacam HR and corneal topography (Kf: r = 0.933/0.930; Ks: r = 0.838/0.829) (P < 0.0001). No significant difference was presented between the 2 investigators for any of the instruments (P = 0.215-0.983). Moreover, high correlation was found between the K readings of the observers (interoperator intraclass correlation coefficients ranged from 0.95 to 0.99).
Conclusions: The Pentacam HR provided reliable K measurements in clinical practice in comparison with an automated keratometer and a corneal topographer. Based on the results, for patient follow-up, one keratometry device is recommended.
Similar articles
-
Comparison of keratometry measurements using the Pentacam HR, the Orbscan IIz, and the TMS-4 topographer.Ophthalmic Physiol Opt. 2012 Nov;32(6):539-46. doi: 10.1111/j.1475-1313.2012.00942.x. Ophthalmic Physiol Opt. 2012. PMID: 23057566
-
Comparability and repeatability of corneal astigmatism measurements using different measurement technologies.J Cataract Refract Surg. 2012 Oct;38(10):1764-70. doi: 10.1016/j.jcrs.2012.05.036. J Cataract Refract Surg. 2012. PMID: 22999600
-
Comparison of the Corneal Power Measurements with the TMS4-Topographer, Pentacam HR, IOL Master, and Javal Keratometer.Middle East Afr J Ophthalmol. 2015 Apr-Jun;22(2):233-7. doi: 10.4103/0974-9233.151884. Middle East Afr J Ophthalmol. 2015. PMID: 25949084 Free PMC article.
-
Comparison and repeatability of keratometric and corneal power measurements obtained by Orbscan II, Pentacam, and Galilei corneal tomography systems.Am J Ophthalmol. 2013 Jul;156(1):53-60. doi: 10.1016/j.ajo.2013.01.029. Epub 2013 Mar 28. Am J Ophthalmol. 2013. PMID: 23540708
-
Keratometry and ultrastructural changes after microwave thermokeratoplasty in rabbit eyes.Lasers Surg Med. 2022 Apr;54(4):565-571. doi: 10.1002/lsm.23495. Epub 2021 Nov 22. Lasers Surg Med. 2022. PMID: 34806777 Review.
Cited by
-
Longitudinal change of refraction over at least 5 years in 15,000 patients.Graefes Arch Clin Exp Ophthalmol. 2013 May;251(5):1431-6. doi: 10.1007/s00417-012-2213-3. Epub 2012 Nov 28. Graefes Arch Clin Exp Ophthalmol. 2013. PMID: 23188521
-
Pentacam® Corneal Tomography for Screening of Refractive Surgery Candidates: A Review of the Literature, Part I.Med Hypothesis Discov Innov Ophthalmol. 2019 Fall;8(3):177-203. Med Hypothesis Discov Innov Ophthalmol. 2019. PMID: 31598520 Free PMC article. Review.
-
Keratometry in normal cats: a cross-sectional study in Japan using an automated handheld keratometer.J Vet Med Sci. 2021 Aug 12;83(8):1256-1262. doi: 10.1292/jvms.21-0104. Epub 2021 Jun 24. J Vet Med Sci. 2021. PMID: 34162774 Free PMC article.
-
Comparison of the reliability of corneal curvature and eccentricity measurements in myopic eyes using four different devices.Front Med (Lausanne). 2025 Apr 4;12:1519992. doi: 10.3389/fmed.2025.1519992. eCollection 2025. Front Med (Lausanne). 2025. PMID: 40255596 Free PMC article.
-
A novel Hartman Shack-based topography system: repeatability and agreement for corneal power with Scheimpflug+Placido topographer and rotating prism auto-keratorefractor.Int Ophthalmol. 2015 Dec;35(6):869-80. doi: 10.1007/s10792-015-0065-7. Epub 2015 Mar 28. Int Ophthalmol. 2015. PMID: 25820517