Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Randomized Controlled Trial
. 2011 Oct;55(9):1090-7.
doi: 10.1111/j.1399-6576.2011.02498.x. Epub 2011 Sep 8.

GlideScope videolaryngoscope vs. Macintosh direct laryngoscope for intubation of morbidly obese patients: a randomized trial

Affiliations
Randomized Controlled Trial

GlideScope videolaryngoscope vs. Macintosh direct laryngoscope for intubation of morbidly obese patients: a randomized trial

L H Andersen et al. Acta Anaesthesiol Scand. 2011 Oct.

Abstract

Background: Morbidly obese patients are at increased risk of hypoxemia during tracheal intubation because of increased frequency of difficult and impossible intubation and a decreased apnea tolerance. In this study, intubation with the GlideScope videolaryngoscope (GS) was compared with the Macintosh direct laryngoscope (DL) in a group of morbidly obese patients.

Methods: One hundred consecutive patients (body mass index ≥ 35 kg/m(2) ) scheduled for bariatric surgery were randomized 1 : 1 to intubation with GS (group GS) or DL (group DL). The primary outcome was intubation time. Secondary outcomes were number of attempts, Cormack-Lehane grade, intubation difficulty scale score (IDS), subjective difficulty of intubation, desaturation, airway bleeding, postoperative sore throat, and hoarseness. Group assignment was not blinded.

Results: Intubation in group GS and group DL lasted 48 (22-148) and 32 s (17-209), respectively (median (range); P = 0.0001); median difference 11 s (95% confidence interval 6-17). Laryngoscopic views were better in group GS with Cormack-Lehane grades 1/2/3/4 distributed as 35/13/2/0 vs. 23/13/10/4 in group DL (P = 0.003). IDS scores were significantly lower with GS than with DL. No other statistically significant differences were found. Two cases of failed intubation occurred in group DL vs. none in group GS (non-significant). Both patients were intubated with the GlideScope without problems.

Conclusion: Intubation of morbidly obese patients with GS was slightly slower than with DL. The increased intubation time was of no clinical consequence as no patients became hypoxemic. Both devices generally performed well in the studied population, but the GS provided better laryngoscopic views and decreased IDS scores.

PubMed Disclaimer

Comment in

Publication types

LinkOut - more resources