Comparison of a 28-channel receive array coil and quadrature volume coil for morphologic imaging and T2 mapping of knee cartilage at 7T
- PMID: 22095723
- PMCID: PMC3265647
- DOI: 10.1002/jmri.23506
Comparison of a 28-channel receive array coil and quadrature volume coil for morphologic imaging and T2 mapping of knee cartilage at 7T
Erratum in
- J Magn Reson Imaging. 2012 Apr;35(4):993
Abstract
Purpose: To compare a new birdcage-transmit, 28-channel receive array (28-Ch) coil and a quadrature volume coil for 7T morphologic MRI and T2 mapping of knee cartilage.
Materials and methods: The right knees of 10 healthy subjects were imaged on a 7T whole body magnetic resonance (MR) scanner using both coils. 3D fast low-angle shot (3D-FLASH) and multiecho spin-echo (MESE) sequences were implemented. Cartilage signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), thickness, and T2 values were assessed.
Results: SNR/CNR was 17%-400% greater for the 28-Ch compared to the quadrature coil (P ≤ 0.005). Bland-Altman plots show mean differences between measurements of tibial/femoral cartilage thickness and T2 values obtained with each coil to be small (-0.002 ± 0.009 cm / 0.003 ± 0.011 cm) and large (-6.8 ± 6.7 msec/-8.2 ± 9.7 msec), respectively. For the 28-Ch coil, when parallel imaging with acceleration factors (AF) 2, 3, and 4 was performed SNR retained was: 62%-69%, 51%-55%, and 39%-45%.
Conclusion: A 28-Ch knee coil provides increased SNR/CNR for 7T cartilage morphologic imaging and T2 mapping. Coils should be switched with caution during clinical studies because T2 values may differ. The greater SNR of the 28-Ch coil could be used to perform parallel imaging with AF2 and obtain similar SNR as the quadrature coil.
Copyright © 2011 Wiley Periodicals, Inc.
Figures
References
-
- Regatte RR, Schweitzer ME. Ultra-high-field MRI of the musculoskeletal system at 7.0T. J Magn Reson Imaging. 2007 Feb;25(2):262–9. - PubMed
-
- Gold GE, Suh B, Sawyer-Glover A, Beaulieu C. Musculoskeletal MRI at 3.0 T: initial clinical experience. AJR Am J Roentgenol. 2004 Nov;183(5):1479–86. - PubMed
-
- Krug R, Stehling C, Kelley DA, Majumdar S, Link TM. Imaging of the musculoskeletal system in vivo using ultra-high field magnetic resonance at 7 T. Invest Radiol. 2009 Sep;44(9):613–8. - PubMed
-
- Roemer PB, Edelstein WA, Hayes CE, Souza SP, Mueller OM. The NMR phased array. Magn Reson Med. 1990 Nov;16(2):192–225. - PubMed
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources
Medical