Randomized Controlled Trial
. 2012 Jan;108(1):80-8.
doi: 10.1093/bja/aer393. Epub 2011 Dec 2.

aepEX monitor for the measurement of hypnotic depth in patients undergoing balanced xenon anaesthesia

Affiliations
Free article
Randomized Controlled Trial

aepEX monitor for the measurement of hypnotic depth in patients undergoing balanced xenon anaesthesia

C Stoppe et al. Br J Anaesth. 2012 Jan.
Free article

Abstract

Background: Previously, we showed a significant difference in the measurements of hypnotic depth by the bispectral index (BIS) and auditory-evoked potentials (AEPs) using the A-line autoregressive index during xenon anaesthesia. In the present study, we evaluate the alternative AEP-based auditory-evoked potential index (aepEX) for the measurement of hypnotic depth in patients undergoing general anaesthesia with xenon.

Methods: Forty-two patients undergoing elective abdominal surgery were enrolled in this controlled, double-blinded, randomized, clinical study. Patients were randomized to receive either xenon (n=21) or sevoflurane anaesthesia (n=21). During anaesthesia, BIS values were recorded simultaneously with the aepEX monitoring. The anaesthetist performing the anaesthesia was blinded to the hypnotic depth monitors. After surgery, the incidence of recalls and awareness was evaluated.

Results: Patients' characteristics such as gender, age, and weight did not differ between the groups. The aepEX and BIS values behaved similarly during anaesthesia. The comparison of aepEX values during xenon and sevoflurane anaesthesia revealed significantly lower aepEX values in the xenon group after 25 min [xenon: 32.9 (4.8) vs sevoflurane: 39.3 (9.0); P=0.008] and after 35 min [xenon: 31.4 (6.6) vs sevoflurane: 37.0 (6.8); P=0.012]. During anaesthesia, aepEX values correlated with the clinical evaluation of depth of anaesthesia (e.g. >20% changes of the baseline arterial pressure or heart rate, spontaneous breathing and/or intolerance of mechanical ventilation, coughing, abdominal pressing, sweating, eye tearing).

Conclusions: We found the aepEX monitor to provide index in the range of adequate depth of xenon anaesthesia, when combined with remifentanil infusion in intubated patients undergoing elective abdominal surgery.