Three-dimensional isotropic T2-weighted cervical MRI at 3T: comparison with two-dimensional T2-weighted sequences
- PMID: 22142499
- DOI: 10.1016/j.crad.2011.06.011
Three-dimensional isotropic T2-weighted cervical MRI at 3T: comparison with two-dimensional T2-weighted sequences
Abstract
Aim: To compare three-dimensional (3D) isotropic T2-weighted magnetic resonance imaging (MRI) sequences and reformation with two-dimensional (2D) T2-weighted sequences regarding image quality of the cervical spine at 3T.
Materials and methods: A phantom study was performed using a water-filled cylinder. The signal-to-noise and image homogeneity were evaluated. Fourteen (n=14) volunteers were examined at 3T using 3D isotropic T2-weighted sagittal and conventional 2D T2-weighted sagittal, axial, and oblique sagittal MRI. Multiplanar reformation (MPR) of the 3D T2-weighted sagittal dataset was performed simultaneously with image evaluation. In addition to artefact assessment, the visibility of anatomical structures in the 3D and 2D sequences was qualitatively assessed by two radiologists independently. Cohen's kappa and Wilcoxon signed rank test were used for the statistical analysis.
Result: The 3D isotropic T2-weighted sequence resulted in the highest signal-to-noise ratio (SNR) and lowest non-uniformity (NU) among the sequences in the phantom study. Quantitative evaluation revealed lower NU values of the cerebrospinal fluid (CSF) and muscles in 2D T2-weighted sagittal sequences compared to the 3D volume isotropic turbo spin-echo acquisition (VISTA) sequence. The other NU values revealed no statistically significant difference between the 2D turbo spin-echo (TSE) and 3D VISTA sequences (0.059<p<0.959). 3D VISTA images showed significantly fewer CSF flow artefacts (p<0.001) and better delineated intradural nerve rootlets (p=0.001) and neural foramina (p=0.016) compared to 2D sequences.
Conclusion: A 3D T2 weighted sequence is superior to conventional 2D sequences for the delineation of intradural nerve rootlets and neural foramina and is less affected by CSF flow artefacts.
Copyright © 2011 The Royal College of Radiologists. Published by Elsevier Ltd. All rights reserved.
Similar articles
-
Clinical utility of optimized three-dimensional T1-, T2-, and T2*-weighted sequences in spinal magnetic resonance imaging.Jpn J Radiol. 2017 Apr;35(4):135-144. doi: 10.1007/s11604-017-0621-3. Epub 2017 Feb 23. Jpn J Radiol. 2017. PMID: 28233194 Review.
-
3D isotropic turbo spin-echo intermediate-weighted sequence with refocusing control in knee imaging: comparison study with 3D isotropic fast-field echo sequence.Acta Radiol. 2011 Dec 1;52(10):1119-24. doi: 10.1258/ar.2011.110328. Acta Radiol. 2011. PMID: 22144425
-
Qualitative and quantitative assessment of wrist MRI at 3.0T: comparison between isotropic 3D turbo spin echo and isotropic 3D fast field echo and 2D turbo spin echo.Acta Radiol. 2013 Apr 1;54(3):284-91. doi: 10.1258/ar.2012.120475. Epub 2013 Jan 14. Acta Radiol. 2013. PMID: 23319720
-
[Comparison of three-dimensional gradient echo, turbo spin echo and steady-state gradient echo sequences in axial MRI examination of the cervical spine].Tani Girisim Radyol. 2003 Dec;9(4):432-8. Tani Girisim Radyol. 2003. PMID: 14730951 Turkish.
-
BLADE Sequences in Transverse T2-weighted MR Imaging of the Cervical Spine. Cut-off for Artefacts?Rofo. 2015 May;36(2):102-8. doi: 10.1055/s-0034-1385179. Epub 2014 Sep 22. Rofo. 2015. PMID: 25912327 Review.
Cited by
-
Comparison between 2D and 3D MEDIC for human cervical spinal cord MRI at 3T.J Med Radiat Sci. 2021 Mar;68(1):4-12. doi: 10.1002/jmrs.433. Epub 2020 Sep 15. J Med Radiat Sci. 2021. PMID: 32931647 Free PMC article.
-
Comparison of three-dimensional isotropic and two-dimensional conventional indirect MR arthrography for the diagnosis of rotator cuff tears.Korean J Radiol. 2014 Nov-Dec;15(6):771-80. doi: 10.3348/kjr.2014.15.6.771. Epub 2014 Nov 7. Korean J Radiol. 2014. PMID: 25469089 Free PMC article.
-
Diagnostic utility of 3D MRI sequences in the assessment of central, recess and foraminal stenoses of the spine: a systematic review.Skeletal Radiol. 2024 Dec;53(12):2575-2584. doi: 10.1007/s00256-024-04689-1. Epub 2024 Apr 27. Skeletal Radiol. 2024. PMID: 38676747 Free PMC article.
-
Diagnosis of Nerve Root Compromise of the Lumbar Spine: Evaluation of the Performance of Three-dimensional Isotropic T2-weighted Turbo Spin-Echo SPACE Sequence at 3T.Korean J Radiol. 2017 Jan-Feb;18(1):249-259. doi: 10.3348/kjr.2017.18.1.249. Epub 2017 Jan 5. Korean J Radiol. 2017. PMID: 28096733 Free PMC article.
-
Clinical utility of optimized three-dimensional T1-, T2-, and T2*-weighted sequences in spinal magnetic resonance imaging.Jpn J Radiol. 2017 Apr;35(4):135-144. doi: 10.1007/s11604-017-0621-3. Epub 2017 Feb 23. Jpn J Radiol. 2017. PMID: 28233194 Review.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Research Materials