Nontraumatic, spontaneous dislocation of polyethylene tibial insert 1 year after TKA
- PMID: 22146212
- DOI: 10.3928/01477447-20111021-25
Nontraumatic, spontaneous dislocation of polyethylene tibial insert 1 year after TKA
Erratum in
- Orthopedics. 2012 Feb;35(2):130. Ravan, Raymond B 3rd [corrected to Raven, Raymond B 3rd]
Abstract
The authors report a case of nontraumatic, spontaneous dislocation of a polyethylene insert detected 1 year after total knee arthroplasty. The patient demonstrated initial improvement and returned to work 4 months postoperatively. At 6 months postoperatively, the patient developed pain and a clunking sensation with motion; however, he denied any traumatic precipitating events. An arthroscopic procedure revealed arthrofibrotic formations but no signs of locking mechanism failure. At 12 months postoperatively, the patient developed sudden instability, and radiographs demonstrated an anteriorly dislodged insert. Revision surgery was performed, and the insert was removed. The insert showed some signs of fatigue due to the locking mechanism. We postulated that repetitive flexion produced an anterior superior force leading to failure of the locking mechanism.
Copyright © 2011, SLACK Incorporated.
Similar articles
-
Dissociation of polyethylene insert from the tibial baseplate following revision total knee arthroplasty.J Arthroplasty. 2011 Feb;26(2):339.e11-3. doi: 10.1016/j.arth.2010.04.016. Epub 2010 May 31. J Arthroplasty. 2011. PMID: 20570104
-
Recurrent dissociation of the tibial insert after mini-subvastus posterior-stabilized total knee arthroplasty: a case report.Knee. 2011 Dec;18(6):461-3. doi: 10.1016/j.knee.2010.08.005. Epub 2010 Sep 15. Knee. 2011. PMID: 20833051
-
Spontaneous dislocation of a mobile-bearing polyethylene insert after posterior-stabilized rotating platform total knee arthroplasty: a case report.Knee. 2011 Dec;18(6):496-8. doi: 10.1016/j.knee.2010.10.007. Epub 2010 Nov 5. Knee. 2011. PMID: 21115353
-
The role of the cemented all-polyethylene tibial component in total knee replacement: a 30-year patient follow-up and review of the literature.Knee. 2010 Dec;17(6):412-6. doi: 10.1016/j.knee.2009.11.008. Epub 2010 Jan 8. Knee. 2010. PMID: 20060725 Review.
-
A step-wise approach to revision TKA.Orthopedics. 2005 Sep;28(9):999-1001. doi: 10.3928/0147-7447-20050901-43. Orthopedics. 2005. PMID: 16190086 Review. No abstract available.
Cited by
-
[Spin out - an avoidable complication in total knee replacement].Orthopade. 2014 Feb;43(2):175-9. doi: 10.1007/s00132-013-2202-1. Orthopade. 2014. PMID: 24154659 German.
-
Spontaneous dislocation of the polyethylene component following knee revision arthroplasty: case report.Rev Bras Ortop. 2014 Dec 29;50(1):114-6. doi: 10.1016/j.rboe.2014.12.002. eCollection 2015 Jan-Feb. Rev Bras Ortop. 2014. PMID: 26229887 Free PMC article.
-
Failure of Polyethelene Insert Locking Mechanism after a Posterior Stabilised Total Knee Arthroplasty- A Case Report.J Orthop Case Rep. 2016 Jul-Aug;6(3):35-37. doi: 10.13107/jocr.2250-0685.492. J Orthop Case Rep. 2016. PMID: 28116264 Free PMC article.