Theoretical estimation of cannulation methods for left ventricular assist device support as a bridge to recovery
- PMID: 22147996
- PMCID: PMC3230019
- DOI: 10.3346/jkms.2011.26.12.1591
Theoretical estimation of cannulation methods for left ventricular assist device support as a bridge to recovery
Abstract
Left ventricular assist device (LVAD) support under cannulation connected from the left atrium to the aorta (LA-AA) is used as a bridge to recovery in heart failure patients because it is non-invasive to ventricular muscle. However, it has serious problems, such as valve stenosis and blood thrombosis due to the low ejection fraction of the ventricle. We theoretically estimated the effect of the in-series cannulation, connected from ascending aorta to descending aorta (AA-DA), on ventricular unloading as an alternative to the LA-AA method. We developed a theoretical model of a LVAD-implanted cardiovascular system that included coronary circulation. Using this model, we compared hemodynamic responses according to various cannulation methods such as LA-AA, AA-DA, and a cannulation connected from the left ventricle to ascending aorta (LV-AA), under continuous and pulsatile LVAD supports. The AA-DA method provided 14% and 18% less left ventricular peak pressure than the LA-AA method under continuous and pulsatile LVAD conditions, respectively. The LA-AA method demonstrated higher coronary flow than AA-DA method. Therefore, the LA-AA method is more advantageous in increasing ventricular unloading whereas the AA-DA method is a better choice to increase coronary perfusion.
Keywords: Bridge to Recovery; Cannulation Methods; Left Ventricular Assist Device (LVAD); Ventricular Unloading.
Figures






References
-
- Bartoli CR, Giridharan GA, Litwak KN, Sobieski M, Prabhu SD, Slaughter MS, Koenig SC. Hemodynamic responses to continuous versus pulsatile mechanical unloading of the failing left ventricle. ASAIO J. 2010;56:410–416. - PubMed
-
- Haithcock BE, Morita H, Fanous NH, Suzuki G, Sabbah HN. Hemodynamic unloading of the failing left ventricle using an arterial-to-arterial extracorporeal flow circuit. Ann Thorac Surg. 2004;77:158–163. - PubMed
-
- Korakianitis T, Shi Y. Numerical comparison of hemodynamics with atrium to aorta and ventricular apex to aorta VAD support. ASAIO J. 2007;53:537–548. - PubMed
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources