Case identification of subjects with airflow limitations using the handheld spirometer "Hi-Checker™" : comparison against an electronic desktop spirometer
- PMID: 22149406
- DOI: 10.3109/15412555.2011.626817
Case identification of subjects with airflow limitations using the handheld spirometer "Hi-Checker™" : comparison against an electronic desktop spirometer
Abstract
Background: Systematic case identification has been proposed as a strategy to improve diagnosis rates and to enable the early detection of subjects with COPD. We hypothesized that case identification could be possible using the handheld spirometer Hi-Checker™.
Aim: To determine how to modify the FEV(1)/FEV(6) values obtained using the Hi-Checker™ to screen for cases with airflow limitation.
Methods: Spirometry was performed with both an electronic desktop spirometer and with the Hi-Checker™ in 312 male subjects.
Results: The average FEV(1) (mean ± SD) measured using a conventional spirometer and the Hi-Checker™ was 2.99 ± 0.56L and 3.07 ± 0.57L, respectively. These results were significantly different (P<0.001, paired t-test for both). This difference of -0.08 ± 0.13L (95% confidence interval: -0.094-0.066L) was normally distributed, and thought to be random. Statistically significant correlations were found for all measurements between the spirometer and the Hi-Checker™ ; the Pearson's correlation coefficient (R) between the FEV(1)/FVC and FEV(1)/FEV(6) values was 0.881. If one defines a FEV(1/)FVC smaller than 0.7 measured by the spirometer as airflow limitation, then a FEV(1)/FEV(6) smaller than 0.746 measured by the Hi-Checker™ best matches this definition, and Cohen's kappa coefficient was 0.672.
Conclusion: Although the Hi-Checker™ estimates resembled those from conventional spirometry, it should be emphasized that the two methods did not produce identical results due to random measurement error. Although one must be careful about overinterpreting these results, since the Hi-Checker™ is small and inexpensive, it could make a significant contribution in facilitating the case selection of patients with airflow limitation.
Similar articles
-
Performance of a new screening spirometer at a community health fair.Respir Care. 2002 Oct;47(10):1150-7. Respir Care. 2002. PMID: 12354333
-
Comparison of a new desktop spirometer (Spirospec) with a laboratory spirometer in a respiratory out-patient clinic.Respir Care. 2003 Jun;48(6):591-5. Respir Care. 2003. PMID: 12780945 Clinical Trial.
-
[Comparison of lung volume response with airflow response to bronchodilator in patients with chronic obstructive pulmonary disease].Zhonghua Jie He He Hu Xi Za Zhi. 2010 Feb;33(2):109-13. Zhonghua Jie He He Hu Xi Za Zhi. 2010. PMID: 20367950 Chinese.
-
Prevalence of airflow limitation among employees in Norwegian smelters: a longitudinal study.Occup Environ Med. 2011 Jan;68(1):24-9. doi: 10.1136/oem.2009.049452. Epub 2010 Aug 26. Occup Environ Med. 2011. PMID: 20798007
-
Lower limit of normal or FEV1/FVC < 0.70 in diagnosing COPD: an evidence-based review.Respir Med. 2011 Jun;105(6):907-15. doi: 10.1016/j.rmed.2011.01.008. Epub 2011 Feb 5. Respir Med. 2011. PMID: 21295958 Review.
Cited by
-
The relationship between the COPD Assessment Test score and airflow limitation in Japan in patients aged over 40 years with a smoking history.Int J Chron Obstruct Pulmon Dis. 2014 Dec 9;9:1357-63. doi: 10.2147/COPD.S61265. eCollection 2014. Int J Chron Obstruct Pulmon Dis. 2014. PMID: 25525353 Free PMC article.
-
The accuracy of a handheld "disposable pneumotachograph device" in the spirometric diagnosis of airway obstruction in a Chinese population.Int J Chron Obstruct Pulmon Dis. 2018 Aug 2;13:2351-2360. doi: 10.2147/COPD.S168583. eCollection 2018. Int J Chron Obstruct Pulmon Dis. 2018. PMID: 30122915 Free PMC article.
-
Detection of airflow limitation using a handheld spirometer in a primary care setting.Respirology. 2014 Jul;19(5):689-93. doi: 10.1111/resp.12291. Epub 2014 Apr 7. Respirology. 2014. PMID: 24708063 Free PMC article.
-
Early Detection of Chronic Obstructive Pulmonary Disease in Primary Care.Intern Med. 2017 Dec 1;56(23):3153-3158. doi: 10.2169/internalmedicine.8717-16. Epub 2017 Sep 25. Intern Med. 2017. PMID: 28943559 Free PMC article.
-
Validity of a portable spirometer in the communities of China.BMC Pulm Med. 2022 Mar 5;22(1):80. doi: 10.1186/s12890-022-01872-9. BMC Pulm Med. 2022. PMID: 35248001 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical