A feasibility study of using couch-based real time dosimetric device in external beam radiotherapy
- PMID: 22149836
- DOI: 10.1118/1.3660773
A feasibility study of using couch-based real time dosimetric device in external beam radiotherapy
Abstract
Purpose: Measurement of actual dose delivered during radiotherapy treatment aids in checking the accuracy of dose delivered to the patient. In this study, a couch-based real time dosimetric device has been proposed to measure the exit or entrance dose to a patient during external beam radiotherapy. The utility and feasibility of such a device using a 2D array of diodes has been demonstrated.
Methods: Two MAPCHECK devices: MAPCHECK (1175) and MAPCHECK 2 (both SunNuclear) were embedded in a foam block in the treatment couch of a Varian 21iX linear accelerator. The angular dependence of the detector response for both devices was studied before implementing the MAPCHECKs for experimental purposes. An Alderson Rando head phantom was scanned with the MAPCHECK and MAPCHECK 2 devices separately and four different treatment plans were generated with target volumes at three different positions simulating typical clinical situations. The analytical anisotropic algorithm (AAA) was used to compute the doses in an Eclipse treatment planning system (Varian Medical Systems). The Rando phantom with the MAPCHECK device was exposed in Clinac 21iX linear accelerator. The measured dose distribution was compared with the calculated dose distribution to check for the accuracy in dose delivery.
Results: Measured and computed dose distribution were found to agree with more than 93% of pixels passing at 3% and 3 mm gamma criteria for all the treatment plans. The couch-based real time dosimetry system may also be applied for noncoplanar beams where electronic portal imaging device (EPID) is not practical to measure the dose. Other advantages include checking the beam stability during the patient treatment, performing routine morning quality assurance (QA) tests in the linear accelerator, and to perform pretreatment verification of intensity modulated radiation therapy (IMRT). One of the drawbacks of this system is that it cannot be used for measuring the dose at 90° or 270° gantry angles.
Conclusions: This preliminary study shows that a 2D array of detectors may be used as part of the treatment couch for real time patient dosimetry in studying the dose delivered to the patient in real time and also for performing routine quality assurance.
Similar articles
-
Evaluation of the accuracy of 3DVH software estimates of dose to virtual ion chamber and film in composite IMRT QA.Med Phys. 2012 Jan;39(1):81-6. doi: 10.1118/1.3666771. Med Phys. 2012. PMID: 22225277
-
Evaluating dosimetric accuracy of flattening filter free compensator-based IMRT: measurements with diode arrays.Med Phys. 2012 Jan;39(1):342-52. doi: 10.1118/1.3671936. Med Phys. 2012. PMID: 22225304
-
MapCHECK used for rotational IMRT measurements: step-and-shoot, TomoTherapy, RapidArc.Med Phys. 2010 Jun;37(6):2837-46. doi: 10.1118/1.3431994. Med Phys. 2010. PMID: 20632595
-
A literature review of electronic portal imaging for radiotherapy dosimetry.Radiother Oncol. 2008 Sep;88(3):289-309. doi: 10.1016/j.radonc.2008.07.008. Epub 2008 Aug 14. Radiother Oncol. 2008. PMID: 18706727 Review.
-
Dosimetric effects caused by couch tops and immobilization devices: report of AAPM Task Group 176.Med Phys. 2014 Jun;41(6):061501. doi: 10.1118/1.4876299. Med Phys. 2014. PMID: 24877795 Review.
Cited by
-
The report of Task Group 100 of the AAPM: Application of risk analysis methods to radiation therapy quality management.Med Phys. 2016 Jul;43(7):4209. doi: 10.1118/1.4947547. Med Phys. 2016. PMID: 27370140 Free PMC article.
-
Real-time dosimetry in external beam radiation therapy.World J Radiol. 2013 Oct 28;5(10):352-5. doi: 10.4329/wjr.v5.i10.352. World J Radiol. 2013. PMID: 24179630 Free PMC article.
-
Noncoplanar verification: a feasibility study using Philips' Pinnacle3 treatment planning system.J Appl Clin Med Phys. 2015 Nov 8;16(6):84–90. doi: 10.1120/jacmp.v16i6.5492. J Appl Clin Med Phys. 2015. PMID: 26699558 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources