Development of a novel quality assurance system based on rolled-up and rolled-out radiochromic films in volumetric modulated arc therapy
- PMID: 22149851
- DOI: 10.1118/1.3659706
Development of a novel quality assurance system based on rolled-up and rolled-out radiochromic films in volumetric modulated arc therapy
Abstract
Purpose: To develop a cylindrical phantom with rolled-up radiochromic films and dose analysis software in the rolled-out plane for quality assurance (QA) in volumetric modulated arc therapy (VMAT).
Methods: The phantom consists of an acrylic cylindrical body wrapped with radiochromic film inserted into an outer cylindrical shell of 5 cm thickness. The rolled-up films with high spatial resolution enable detection of specific dose errors along the arc trajectory of continuously irradiated and modulated beams in VMAT. The developed dose analysis software facilitates dosimetric evaluation in the rolled-up and rolled-out planes of the film; the calculated doses on the corresponding points where the rolled-up film was placed were reconstructed into a rectangular dose matrix equivalent to that of the rolled-out plane of the film. The VMAT QA system was implemented in 3 clinical cases of prostate, nasopharynx, and pelvic metastasis. Each calculated dose on the rolled-out plane was compared with measurement values by modified gamma evaluation. Detected positions of dose disagreement on the rolled-out plane were also distinguished in cylindrical coordinates. The frequency of error occurrence and error distribution were summarized in a histogram and in an axial view of rolled-up plane to intuitively identify the corresponding positions of detected errors according to the gantry angle.
Results: The dose matrix reconstructed from the developed VMAT QA system was used to verify the measured dose distribution along the arc trajectory. Dose discrepancies were detected on the rolled-out plane and visualized on the calculated dose matrix in cylindrical coordinates. The error histogram obtained by gamma evaluation enabled identification of the specific error frequency at each gantry angular position. The total dose error occurring on the cylindrical surface was in the range of 5%-8% for the 3 cases.
Conclusions: The developed system provides a practical and reliable QA method to detect dosimetric errors according to the gantry angle. Film dosimetry based on rolled-up and rolled-out techniques leads to dose verification in the subspaces of the 3D dose volume. The system can be employed as an alternative tool to detect the pitfalls of planar dose verification.
Similar articles
-
Three-dimensional radiochromic film dosimetry for volumetric modulated arc therapy using a spiral water phantom.J Radiat Res. 2013 Nov 1;54(6):1153-9. doi: 10.1093/jrr/rrt059. Epub 2013 May 17. J Radiat Res. 2013. PMID: 23685667 Free PMC article.
-
Novel dosimetric phantom for quality assurance of volumetric modulated arc therapy.Med Phys. 2009 May;36(5):1813-21. doi: 10.1118/1.3117563. Med Phys. 2009. PMID: 19544800
-
An oscillating sweeping gap test for VMAT quality assurance.Phys Med Biol. 2010 Sep 7;55(17):5029-44. doi: 10.1088/0031-9155/55/17/010. Epub 2010 Aug 11. Phys Med Biol. 2010. PMID: 20702923
-
Evaluation of the accuracy of 3DVH software estimates of dose to virtual ion chamber and film in composite IMRT QA.Med Phys. 2012 Jan;39(1):81-6. doi: 10.1118/1.3666771. Med Phys. 2012. PMID: 22225277
-
Error detection model developed using a multi-task convolutional neural network in patient-specific quality assurance for volumetric-modulated arc therapy.Med Phys. 2021 Sep;48(9):4769-4783. doi: 10.1002/mp.15031. Epub 2021 Jul 29. Med Phys. 2021. PMID: 34101848
Cited by
-
Are simple IMRT beams more robust against MLC error? Exploring the impact of MLC errors on planar quality assurance and plan quality for different complexity beams.J Appl Clin Med Phys. 2016 May 8;17(3):147-157. doi: 10.1120/jacmp.v17i3.6022. J Appl Clin Med Phys. 2016. PMID: 27167272 Free PMC article.
-
IMRT dose verification considering passing rate and respiratory motion.Oncol Lett. 2018 Jul;16(1):963-969. doi: 10.3892/ol.2018.8724. Epub 2018 May 16. Oncol Lett. 2018. PMID: 29963170 Free PMC article.
-
Three-dimensional radiochromic film dosimetry for volumetric modulated arc therapy using a spiral water phantom.J Radiat Res. 2013 Nov 1;54(6):1153-9. doi: 10.1093/jrr/rrt059. Epub 2013 May 17. J Radiat Res. 2013. PMID: 23685667 Free PMC article.
-
A study on the correlation between radiation field size and gamma index passing rate for MatriXX.Medicine (Baltimore). 2019 Jul;98(30):e16536. doi: 10.1097/MD.0000000000016536. Medicine (Baltimore). 2019. PMID: 31348271 Free PMC article.
-
Investigation on the effect of sharp phantom edges on point dose measurement during patient-specific dosimetry with Rapid Arc.J Med Phys. 2013 Jul;38(3):139-42. doi: 10.4103/0971-6203.116373. J Med Phys. 2013. PMID: 24049321 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources