Wireless instrumentation system based on dry electrodes for acquiring EEG signals
Abstract
This paper presents a complete non-invasive Wireless acquisition system based on dry electrodes for electroencephalograms (WiDE-EEG) with emphasis in the electronic system design. The WiDE-EEG is composed by a 2.4 GHz radio-frequency (RF) transceiver, biopotential acquisition electronics and dry electrodes. The WiDE-EEG can acquire electroencephalogram (EEG) signals from 5 unipolar channels, with a resolution of 16 bits and minimum analog amplitude of 9.98 μV(pp), at a sampling rate of 1000 samples/s/channel and sends them to a processing unit through RF in a 10 m range. The analog channels were optimized for EEG signals (with amplitudes in the range 70-100 μV) and present the following characteristics: a signal gain of 66 dB and a common mode rejection ratio of 56.5 dB. Each electrode is composed by 16 microtip structures that were fabricated through bulk micromachining of a <100>-type silicon substrate in a potassium hydroxide (KOH) solution. The microtips present solid angles of 54.7°, a height of 100-200 μm and 2 μm spaced apart. The electrodes have a thin layer (obtained by sputtering) of iridium oxide (IrO) to guaranty their biocompatibility and improve the contact with the skin. These dry electrodes are in direct contact with the electrolyte fluids of the inner skin layers, and avoid the use of conductive gels. The complete WiDE-EEG occupies a volume of 9 cm×8.5 cm×1 cm, which makes it suitable for true mobility of the subjects and at the same time allows high data transfer rates. Since the WiDE-EEG is battery-powered, it overcomes the need of galvanic isolation for ensuring patient safety observed on conventional EEG instrumentation systems. The WiDE-EEG presents a total power consumption of 107 mW, divided as follows: the acquisition system contributes with 10 mW per channel, whereas the commercial MICAz module contributes with 57 mW (e.g., 24 mW from the microcontroller and 33 mW from the RF chip). The WiDE-EEG also presents autonomy of about 25 h with two class AA 1.5 V batteries.
Copyright © 2011 IPEM. Published by Elsevier Ltd. All rights reserved.
Similar articles
-
Dry and noncontact EEG sensors for mobile brain-computer interfaces.IEEE Trans Neural Syst Rehabil Eng. 2012 Mar;20(2):228-35. doi: 10.1109/TNSRE.2011.2174652. Epub 2011 Dec 12. IEEE Trans Neural Syst Rehabil Eng. 2012. PMID: 22180514
-
Wireless recording systems: from noninvasive EEG-NIRS to invasive EEG devices.IEEE Trans Biomed Circuits Syst. 2013 Apr;7(2):186-95. doi: 10.1109/TBCAS.2013.2255595. IEEE Trans Biomed Circuits Syst. 2013. PMID: 23853301
-
Multichannel PC-based data-acquisition system for high-resolution EEG.IEEE Trans Biomed Eng. 1995 Dec;42(12):1212-7. doi: 10.1109/10.476129. IEEE Trans Biomed Eng. 1995. PMID: 8550064
-
Active Electrodes for Wearable EEG Acquisition: Review and Electronics Design Methodology.IEEE Rev Biomed Eng. 2017;10:187-198. doi: 10.1109/RBME.2017.2656388. Epub 2017 Jan 20. IEEE Rev Biomed Eng. 2017. PMID: 28113349 Review.
-
Dry-contact and noncontact biopotential electrodes: methodological review.IEEE Rev Biomed Eng. 2010;3:106-19. doi: 10.1109/RBME.2010.2084078. IEEE Rev Biomed Eng. 2010. PMID: 22275204 Review.
Cited by
-
Research on fatigue detection of flight trainees based on face EMF feature model combination with PSO-CNN algorithm.Sci Rep. 2024 Sep 4;14(1):20641. doi: 10.1038/s41598-024-71192-x. Sci Rep. 2024. PMID: 39232069 Free PMC article.
-
Acquisition technology research of EEG and related physiological signals under +Gz acceleration.Ir J Med Sci. 2014 Jun;183(2):187-97. doi: 10.1007/s11845-013-0988-9. Epub 2013 Jul 17. Ir J Med Sci. 2014. PMID: 23860826
-
Neurobehavioural signatures in race car driving: a case study.Sci Rep. 2020 Jul 14;10(1):11537. doi: 10.1038/s41598-020-68423-2. Sci Rep. 2020. PMID: 32665679 Free PMC article.
-
Dry Electrodes for Human Bioelectrical Signal Monitoring.Sensors (Basel). 2020 Jun 29;20(13):3651. doi: 10.3390/s20133651. Sensors (Basel). 2020. PMID: 32610658 Free PMC article. Review.
-
Simulation of an optimized technique based on DS-CDMA for simultaneous transmission of multichannel biosignals.Biomed Eng Lett. 2017 Feb 28;7(2):153-171. doi: 10.1007/s13534-017-0018-3. eCollection 2017 May. Biomed Eng Lett. 2017. PMID: 30603162 Free PMC article.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Miscellaneous