Optimizing venous drainage using an ultrasonic flow probe on the venous line
- PMID: 22164455
- PMCID: PMC4679976
Optimizing venous drainage using an ultrasonic flow probe on the venous line
Abstract
The use of smaller cannulae for minimally invasive surgery techniques and/or aggressive miniaturization of the cardiopulmonary bypass (CPB) circuitry has necessitated the need to augment venous drainage to achieve adequate flow rates. Vacuum assisted venous drainage (VAVD) has become the dominant method to augment venous drainage. VAVD, however, has been associated with a number of known side effects including increased transmission of gaseous microemboli to the patient, venous line chatter, and increased arterial to venous shunts in the circuit. Historically, our practice has been to monitor the arterial output flow rate and to monitor VAVD by observing venous line chatter and changes in the venous reservoir level. In 2008 our pediatric cardiothoracic service began monitoring venous line flow rates by using a second ultrasonic flow probe placed on the venous line. After 12 months, our staff perfusionists reviewed the impact of monitoring venous line flow rates on VAVD and its known side effects on daily clinical practice. When monitoring venous line flow rates, empiric observation revealed that less overall vacuum pressure was needed for our CPB cases. This novel approach to monitoring venous drainage has aided us in providing optimal vacuum levels and therefore, may reduce some of the known side effects experienced with excessive VAVD.
Conflict of interest statement
The senior author has stated that authors have reported no material, financial, or other relationship with any healthcare-related business or other entity whose products or services are discussed in this paper.
Figures
References
-
- Colangelo N, Torracca L, Lapenna E, Moriggia S, Crescenzi G, Alfieri O.. Vacuum-assisted venous drainage in extrathoracic cardiopulmonary bypass management during minimally invasive cardiac surgery. Perfusion. 2006;21:361–5. - PubMed
-
- Bonatti J, Schachner T, Bonaros N, Laufer G.. A new exposure technique for the circumflex coronary artery system in robotic totally endoscopic coronary artery bypass grafting. Interact Cardiovasc Thorac Surg. 2006;5:279–81. - PubMed
-
- Gunaydin S, Sari T, McCusker K, Schonrock U, Zorlutuna Y.. Clinical evaluation of minimized extracorporeal circulation in high-risk coronary revascularization: Impact on air handling, inflammation, hemodilution and myocardial function. Perfusion. 2009;24:153–62. - PubMed
-
- Pappalardo F, Corno C, Franco A, et al. Reduction of hemodilution in small adults undergoing open heart surgery: A prospective, randomized trial. Perfusion. 2007;22:317–22. - PubMed
-
- Immer FF, Ackermann A, Gygax E, et al. Minimal extracorporeal circulation is a promising technique for coronary artery bypass grafting. Ann Thorac Surg. 2007;84:1515–20; discussion 1521. - PubMed
MeSH terms
LinkOut - more resources
Full Text Sources