Simultaneous analysis of cortisol and cortisone in saliva using XLC-MS/MS for fully automated online solid phase extraction
- PMID: 22178191
- DOI: 10.1016/j.jchromb.2011.11.036
Simultaneous analysis of cortisol and cortisone in saliva using XLC-MS/MS for fully automated online solid phase extraction
Abstract
Salivary cortisol measurements are increasingly being used in the investigation of disorders of the hypothalamic-pituitary-adrenal axis. In the salivary gland, cortisol is metabolised to cortisone by the action of 11β-hydroxysteroid dehydrogenase type 2, and cortisone is partly responsible for the variable interference observed in current salivary cortisol immunoassays. The aim of this study was to validate an assay for the simultaneous analysis of salivary cortisol and cortisone using the Spark Holland Symbiosis™ in eXtraction liquid chromatography-tandem mass spectrometry (XLC-MS/MS) mode for fully automated online solid phase extraction (SPE). Saliva samples were diluted in water with the addition of internal standard (d4-cortisol and d7-cortisone). Online SPE was performed using the Spark Holland Symbiosis™ with HySphere™ C18 SPE cartridges and compounds were eluted onto a Phenomenex® C18 guard column attached to a Phenomenex® Onyx monolithic C18 column for chromatography. Mass spectrometry used the Waters® Xevo™ TQ MS in electrospray positive mode. Cortisol and cortisone eluted with their internal standards at 1.95 and 2.17 min, respectively, with a total run time of four minutes. No evidence of ion-suppression was observed. The assay was linear up to 3393 nmol/L for cortisol and 3676 nmol/L for cortisone, with lower limits of quantitation of 0.75 nmol/L and 0.50 nmol/L, respectively. Intra- and inter-assay imprecision was <8.9% for cortisol and <6.5% for cortisone across three levels of internal quality control, with accuracy and recovery within accepted limits. High specificity was demonstrated following interference studies which assessed 29 structurally-related steroids at supra-physiological concentrations. We have successfully validated an assay for the simultaneous analysis of salivary cortisol and cortisone using XLC-MS/MS and fully automated online SPE. The assay benefits from increased specificity compared to immunoassay and minimal sample preparation which allows high sample throughput and is thus suitable for use in a routine clinical laboratory.
Copyright © 2011 Elsevier B.V. All rights reserved.
Similar articles
-
Simultaneous measurement of cortisol and cortisone in human saliva using liquid chromatography-tandem mass spectrometry: application in basal and stimulated conditions.J Chromatogr B Analyt Technol Biomed Life Sci. 2009 Nov 1;877(29):3771-5. doi: 10.1016/j.jchromb.2009.09.014. Epub 2009 Sep 18. J Chromatogr B Analyt Technol Biomed Life Sci. 2009. PMID: 19783236
-
A simplified liquid chromatography tandem mass spectrometry assay, using on-line solid-phase extraction, for the quantitation of cortisol in saliva and comparison with a routine DELFIA method.Ann Clin Biochem. 2010 Mar;47(Pt 2):131-6. doi: 10.1258/acb.2009.009053. Epub 2010 Feb 11. Ann Clin Biochem. 2010. PMID: 20150214
-
Simultaneous measurements of cortisol and cortisone in urine and hair for the assessment of 11β-hydroxysteroid dehydrogenase activity among methadone maintenance treatment patients with LC-ESI-MS/MS.J Chromatogr B Analyt Technol Biomed Life Sci. 2014 Oct 15;969:77-84. doi: 10.1016/j.jchromb.2014.08.001. Epub 2014 Aug 10. J Chromatogr B Analyt Technol Biomed Life Sci. 2014. PMID: 25151333
-
Novel liquid chromatography tandem mass spectrometry (LC-MS/MS) methods for measuring steroids.Best Pract Res Clin Endocrinol Metab. 2013 Oct;27(5):663-74. doi: 10.1016/j.beem.2013.05.015. Epub 2013 Jul 10. Best Pract Res Clin Endocrinol Metab. 2013. PMID: 24094637 Review.
-
Therapeutic drug monitoring and LC-MS/MS.J Chromatogr B Analyt Technol Biomed Life Sci. 2012 Feb 1;883-884:33-49. doi: 10.1016/j.jchromb.2011.09.041. Epub 2011 Sep 29. J Chromatogr B Analyt Technol Biomed Life Sci. 2012. PMID: 21992751 Review.
Cited by
-
The Impact of Goal Disturbance after Cancer on Cortisol Levels over Time and the Moderating Role of COMT.PLoS One. 2015 Aug 27;10(8):e0135708. doi: 10.1371/journal.pone.0135708. eCollection 2015. PLoS One. 2015. PMID: 26313260 Free PMC article.
-
Bioavailability of Oral Hydrocortisone Corrected for Binding Proteins and Measured by LC-MS/MS Using Serum Cortisol and Salivary Cortisone.J Bioequivalence Bioavailab. 2018;10(1):001-3. doi: 10.4172/jbb.1000365. J Bioequivalence Bioavailab. 2018. PMID: 29795974 Free PMC article.
-
Morning Salivary Cortisone Versus Serum Cortisol in the Overnight Dexamethasone Suppression Test (ODNST): Evaluation in a Clinical Setting.Clin Endocrinol (Oxf). 2025 Jul;103(1):13-20. doi: 10.1111/cen.15233. Epub 2025 Mar 21. Clin Endocrinol (Oxf). 2025. PMID: 40114586 Free PMC article.
-
Current methods for stress marker detection in saliva.J Pharm Biomed Anal. 2020 Nov 30;191:113604. doi: 10.1016/j.jpba.2020.113604. Epub 2020 Sep 6. J Pharm Biomed Anal. 2020. PMID: 32957066 Free PMC article. Review.
-
Waking Salivary Cortisone vs Serum Cortisol in the Short Synacthen Test in Screening for Adrenocortical Insufficiency: Results of a Service Evaluation.Clin Endocrinol (Oxf). 2025 Sep;103(3):396-400. doi: 10.1111/cen.15279. Epub 2025 May 27. Clin Endocrinol (Oxf). 2025. PMID: 40432285 Free PMC article. No abstract available.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources