Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Comparative Study
. 2012 Oct;19(5):611-6.
doi: 10.1016/j.knee.2011.11.004. Epub 2011 Dec 16.

Computer guided restoration of joint line and femoral offset in cruciate substituting total knee arthroplasty

Affiliations
Comparative Study

Computer guided restoration of joint line and femoral offset in cruciate substituting total knee arthroplasty

Gautam M Shetty et al. Knee. 2012 Oct.

Abstract

Purpose: This prospective study aimed to evaluate radiographically, change in joint line and femoral condylar offset with the optimized gap balancing technique in computer-assisted, primary, cruciate-substituting total knee arthroplasties (TKAs).

Methods: One hundred and twenty-nine consecutive computer-assisted TKAs were evaluated radiographically using pre- and postoperative full-length standing hip-to-ankle, antero-posterior and lateral radiographs to assess change in knee deformity, joint line height and posterior condylar offset.

Results: In 49% of knees, there was a net decrease (mean 2.2mm, range 0.2-8.4mm) in joint line height postoperatively whereas 46.5% of knees had a net increase in joint line height (mean 2.5mm, range 0.2-11.2mm). In 93% of the knees, joint line was restored to within ± 5 mm of preoperative values. In 53% of knees, there was a net increase (mean 2.9 mm, range 0.2-12 mm) in posterior offset postoperatively whereas 40% of knees had a net decrease in posterior offset (mean 4.2mm, range 0.6-20mm). In 82% of knees, the posterior offset was restored within ± 5 mm of preoperative values.

Conclusions: Based on radiographic evaluation in extension and at 30° flexion, the current study clearly demonstrates that joint line and posterior femoral condylar offset can be restored in the majority of computer-assisted, cruciate-substituting TKAs to within 5mm of their preoperative value. The optimized gap balancing feature of the computer software allows the surgeon to simulate the effect of simultaneously adjusting femoral component size, position and distal femoral resection level on joint line and posterior femoral offset.

PubMed Disclaimer

Publication types

MeSH terms

LinkOut - more resources