Randomized Controlled Trial
. 2012 Sep;79(9):1163-8.
doi: 10.1007/s12098-011-0651-2. Epub 2011 Dec 23.

Bubble vs conventional continuous positive airway pressure for prevention of extubation failure in preterm very low birth weight infants: a pilot study

Affiliations
Randomized Controlled Trial

Bubble vs conventional continuous positive airway pressure for prevention of extubation failure in preterm very low birth weight infants: a pilot study

Sucheta Yadav et al. Indian J Pediatr. 2012 Sep.

Abstract

Objective: To compare the efficacy of bubble and conventional nasal continuous positive airway pressure (CPAP) in preventing extubation failure (EF) in preterm infants.

Methods: Infants of gestation ≤32 wk and birth weight <1500 g, ready for initial extubation within first wk of life were randomly allocated to either bubble (n = 16) or conventional (n = 16) CPAP. A standardized protocol was used for extubation. Bubble CPAP was delivered by Fischer and Paykel equipment using short binasal prongs and conventional CPAP was delivered by a ventilator using Argyle short binasal prongs. CPAP was initiated at a pressure of 4-6 cm of H(2)O and FiO(2) of 0.4-0.5 and adjusted to maintain normal saturation (90-93%) and comfortable breathing. Primary outcome was EF, defined as need for mechanical ventilation within 72 h of extubation.

Results: Baseline characteristics including birth weight (g; 1027 ± 243 vs. 1018 ± 227; p = 0.83), gestation (wk; 28.7 ± 1.8 vs. 28.4 ± 1.6; p = 0.30), infants <28 wk gestation (6 vs. 7, p = 0.72) were comparable between the two groups. Respiratory distress syndrome was the indication for ventilation in 13 (81%) and 14 (87%) infants on bubble CPAP and conventional CPAP groups respectively. (p = 0.99). There was no difference in the EF rates between the bubble (n = 4) and conventional CPAP (n = 9) groups. (RR 0.49; 95% CI 0.20-1.2; p = 0.14). Median time to extubation failure was also comparable between the two groups (h; median [range]: 29 [14-49] vs. 17 [7-28]; p = 0.35).

Conclusions: The possibility that bubble CPAP may be associated with reduced EF as suggested in this pilot study requires further investigation in an adequately powered multicentric study.

PubMed Disclaimer

References

    1. Eur J Pediatr. 1992 Jun;151(6):451-4 - PubMed
    1. Pediatrics. 1956 Jan;17(1):1-10 - PubMed
    1. J Med Eng Technol. 2009;33(3):214-22 - PubMed
    1. J Pediatr. 2009 May;154(5):645-50 - PubMed
    1. Arch Ophthalmol. 2003 Dec;121(12):1684-94 - PubMed

Publication types

Supplementary concepts

LinkOut - more resources