Hemolysis, lipaemia and icterus in specimens for arterial blood gas analysis
- PMID: 22198577
- DOI: 10.1016/j.clinbiochem.2011.12.005
Hemolysis, lipaemia and icterus in specimens for arterial blood gas analysis
Abstract
Objective: To assess prevalence of interference from hemolysis, lipaemia, icterus in arterial blood gas analysis (ABG).
Design: Serum indices (SI) were assessed in the plasma of ABG samples over a 2-month period.
Results: Out of a total of 478 ABG specimens, we identified 17 hemolyzed samples (4%), 52 (11%) with lipaemia, and 63 (13%) with icterus.
Conclusion: Test results on a considerable number of ABG specimens might be unreliable due to presence of interference.
Copyright © 2011 The Canadian Society of Clinical Chemists. Published by Elsevier Inc. All rights reserved.
Similar articles
-
Analysis of hemolysis, icterus and lipemia in arterial blood gas specimens.Clin Chem Lab Med. 2017 Mar 1;55(3):e69-e71. doi: 10.1515/cclm-2016-0656. Clin Chem Lab Med. 2017. PMID: 27559693 No abstract available.
-
Serum indices: managing assay interference.Ann Clin Biochem. 2016 Sep;53(Pt 5):527-38. doi: 10.1177/0004563216643557. Epub 2016 May 3. Ann Clin Biochem. 2016. PMID: 27147624 Review.
-
The incidence rate and influence factors of hemolysis, lipemia, icterus in fasting serum biochemistry specimens.PLoS One. 2022 Jan 19;17(1):e0262748. doi: 10.1371/journal.pone.0262748. eCollection 2022. PLoS One. 2022. PMID: 35045128 Free PMC article.
-
Effect of hemolysis, icterus, and lipemia on three acetaminophen assays: Potential medical consequences of false positive results.Clin Chim Acta. 2018 Dec;487:287-292. doi: 10.1016/j.cca.2018.10.019. Epub 2018 Oct 11. Clin Chim Acta. 2018. PMID: 30317020
-
Hemolyzed specimens: a major challenge for emergency departments and clinical laboratories.Crit Rev Clin Lab Sci. 2011 May-Jun;48(3):143-53. doi: 10.3109/10408363.2011.600228. Crit Rev Clin Lab Sci. 2011. PMID: 21875312 Review.
Cited by
-
Novel Opportunities for Improving the Quality of Preanalytical Phase. A Glimpse to the Future?J Med Biochem. 2017 Oct 28;36(4):293-300. doi: 10.1515/jomb-2017-0029. eCollection 2017 Oct. J Med Biochem. 2017. PMID: 30581325 Free PMC article.
-
Analytical validation of hemolysis detection on GEM Premier 7000.Clin Chem Lab Med. 2025 May 22;63(9):e209-e211. doi: 10.1515/cclm-2025-0525. Print 2025 Aug 26. Clin Chem Lab Med. 2025. PMID: 40418775 No abstract available.
-
Effect of exogenous lipids contamination on blood gas analysis.Adv Lab Med. 2024 May 7;5(3):301-306. doi: 10.1515/almed-2024-0043. eCollection 2024 Sep. Adv Lab Med. 2024. PMID: 39252807 Free PMC article.
-
Frequency and causes of lipemia interference of clinical chemistry laboratory tests.Pract Lab Med. 2017 Feb 3;8:1-9. doi: 10.1016/j.plabm.2017.02.001. eCollection 2017 Aug. Pract Lab Med. 2017. PMID: 28856220 Free PMC article.
-
Blood gas testing and related measurements: National recommendations on behalf of the Croatian Society of Medical Biochemistry and Laboratory Medicine.Biochem Med (Zagreb). 2016 Oct 15;26(3):318-336. doi: 10.11613/BM.2016.036. Biochem Med (Zagreb). 2016. PMID: 27812301 Free PMC article. Review.
MeSH terms
LinkOut - more resources
Full Text Sources