Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Clinical Trial
. 2011 Nov 1;5(6):1602-9.
doi: 10.1177/193229681100500638.

Precision, accuracy, and user acceptance of the OneTouch SelectSimple blood glucose monitoring system

Affiliations
Clinical Trial

Precision, accuracy, and user acceptance of the OneTouch SelectSimple blood glucose monitoring system

Athena Philis-Tsimikas et al. J Diabetes Sci Technol. .

Abstract

Background: The OneTouch® SelectSimple™ blood glucose monitoring system (BGMS) is a device for self-monitoring of blood glucose designed for ease of use. Alarms alert subjects to low [20-69 mg/dl (1.1-3.8 mmol/liter)], high [180-239 mg/dl (9.9-13.2 mmol/liter)], and very high [240-600 mg/dl (13.3-33.1 mmol/liter)] blood glucose readings.

Methods: Repeatability in blood and intermediate precision with aqueous controls were examined using blood from one donor adjusted to different glucose concentrations, and tested with 10 meters and 1 test-strip lot. System accuracy was evaluated with blood samples from 100 diabetes patients tested on 3 test-strip lots, compared with a reference system (YSI 2300 STAT). To test user accuracy, patients (n = 156) and health care professionals (HCPs) tested subject blood with the SelectSimple twice. Health care professionals evaluated subject BGMS technique after a 3-5 day home-testing period. Users evaluated the instructions for use and responded to a user acceptance questionnaire.

Results: In repeatability and intermediate precision testing, the SelectSimple BGMS had a coefficient of variation of ≤ 5% or standard deviation of ≤ 5 mg/dl. In the clinical accuracy study, 100% of measurements <75 mg/dl (4.2 mmol/liter) were within ± 15 mg/dl (0.8 mmol/liter) of reference value, and 99.6% of measurements ≥ 75 mg/dl (4.2 mmol/liter) were within ±20%. Patients were able to use the BGMS appropriately and evaluated it as easy to use. Acceptance of the SelectSimple BGMS was within predefined limits.

Conclusions: In these studies, the SelectSimple BGMS met all criteria for precision, system, and user accuracy, was easy to use, and was well accepted by patients.

PubMed Disclaimer

Figures

Figure 1
Figure 1
OneTouch SelectSimple BGMS.
Figure 2
Figure 2
System accuracy of OneTouch SelectSimple using three different test-strip batches: (A) lot 3022344; (B) lot 3022345; and (C) lot 3043510.
Figure 3
Figure 3
Bias plots for: (A) subject self-test and (B) HCP test data. Solid lines represent the boundaries of ISO accuracy criteria; specifically, ±15 mg/dl for glucose values <75 mg/dl (4.2 mmol/liter), and within ±20% for glucose values ≥75 mg/dl (4.2 mmol/liter) of the laboratory reference YSI 2300.
Figure 4
Figure 4
Percentage difference between OneTouch SelectSimple BGMS and YSI 2300 reference values when measured by subjects and HCPs for reference values: (A) <75 mg/dl (4.2 mmol/liter) and (B) ≥75 mg/dl (4.2 mmol/liter). Y-axis shows the proportion of BGMS results falling within the range denoted by the x-axis.
Figure 5
Figure 5
Consensus error grid showing accuracy of OneTouch SelectSimple test results compared with YSI 2300 reference values for: (A) 300 self-tests performed by subjects and (B) 294 tests performed by HCPs.

Comment in

References

    1. International Diabetes Federation. IDF Diabetes Atlas. Diabetes and Impaired Glucose Tolerance: Global Burden: Prevalence and Projections, 2010 and 2030. http://www.diabetesatlas.org/content/diabetes-and-impaired-glucose-toler.... Accessed March 24, 2011.
    1. International Diabetes Federation. IDF Diabetes Atlas. Morbidity and Mortality. http://www.diabetesatlas.org/content/diabetes-mortality. Accessed March 24, 2011.
    1. Chan JC, Malik V, Jia W, Kadowaki T, Yajnik CS, Yoon KH, Hu FB. Diabetes in Asia: epidemiology, risk factors, and pathophysiology. JAMA. 2009;301(20):2129–2140. - PubMed
    1. Gupta R, Kumar P. Global diabetes landscape—type 2 diabetes mellitus in South Asia: epidemiology, risk factors, and control. Insulin. 2008;3(2):78–94.
    1. Pan XR, Yang WY, Li GW, Liu J. Prevalence of diabetes and its risk factors in China, 1994. National Diabetes Prevention and Control Cooperative Group. Diabetes Care. 1997;20(11):1664–1669. - PubMed

Publication types