Dual- versus single-coil implantable defibrillator leads: review of the literature
- PMID: 22231644
- DOI: 10.1007/s00392-011-0407-z
Dual- versus single-coil implantable defibrillator leads: review of the literature
Abstract
The preferred use of dual-coil implantable defibrillator lead systems in current implantable defibrillator therapy is likely based on data showing statistically lower defibrillation thresholds with dual-coil defibrillator lead systems. The following review will summarize the clinical data for dual- versus single-coil defibrillator leads in the left and right pectoral implant locations, and will then discuss the clinical implications of single- versus dual-coil usage for atrial defibrillation, venous complications, and the risks associated with lead extraction. It will be noted that there are no comparative clinical studies on the use and outcomes of single- versus dual-coil lead systems in implantable defibrillator therapy over a long-term follow-up. The limited long-term reliability of defibrillator leads is a major concern in implantable defibrillator and cardiac resynchronization therapy. A simpler single-coil defibrillator lead system may improve the long-term performance of implanted leads. Furthermore, the superior vena cava coil is suspected to increase interventional risk in transvenous lead extraction. Therefore, the need for objective data on extractions and complications will be emphasized.
Similar articles
-
Comparison of single-coil lead versus dual-coil lead of implantable cardioverter defibrillator on lead-related venous complications in a canine model.J Interv Card Electrophysiol. 2018 Jul;52(2):195-201. doi: 10.1007/s10840-018-0312-8. Epub 2018 Mar 23. J Interv Card Electrophysiol. 2018. PMID: 29572716
-
Effectiveness and safety of transvenous extraction of single- versus dual-coil implantable cardioverter-defibrillator leads at single-center experience.Medicine (Baltimore). 2019 Jul;98(30):e16548. doi: 10.1097/MD.0000000000016548. Medicine (Baltimore). 2019. PMID: 31348275 Free PMC article.
-
Clinical predictors of atrial defibrillation thresholds with a dual-coil, active pectoral lead system.Heart Rhythm. 2005 Jan;2(1):49-54. doi: 10.1016/j.hrthm.2004.10.033. Heart Rhythm. 2005. PMID: 15851265
-
Should Single-Coil Implantable Cardioverter Defibrillator Leads Be Used in all Patients?Card Electrophysiol Clin. 2018 Mar;10(1):59-66. doi: 10.1016/j.ccep.2017.11.010. Card Electrophysiol Clin. 2018. PMID: 29428142 Review.
-
[Atrial defibrillator].Herz. 1998 Jun;23(4):251-9. doi: 10.1007/BF03044320. Herz. 1998. PMID: 9690112 Review. German.
Cited by
-
Possible spontaneous PFO closure after thrombus trapped in PFO.Clin Res Cardiol. 2014 Apr;103(4):333-5. doi: 10.1007/s00392-013-0655-1. Epub 2014 Jan 19. Clin Res Cardiol. 2014. PMID: 24442263 No abstract available.
-
Postcardiac injury syndrome after cardiac implantable electronic device implantation.Herz. 2020 Nov;45(7):696-702. doi: 10.1007/s00059-020-04910-6. Epub 2020 Mar 13. Herz. 2020. PMID: 32170340 Free PMC article.
-
Low efficacy of cardioversion of persistent atrial fibrillation with the implantable cardioverter-defibrillator.Neth Heart J. 2013 Dec;21(12):548-53. doi: 10.1007/s12471-013-0474-z. Neth Heart J. 2013. PMID: 24092363 Free PMC article.
-
Comparison of single-coil lead versus dual-coil lead of implantable cardioverter defibrillator on lead-related venous complications in a canine model.J Interv Card Electrophysiol. 2018 Jul;52(2):195-201. doi: 10.1007/s10840-018-0312-8. Epub 2018 Mar 23. J Interv Card Electrophysiol. 2018. PMID: 29572716
-
Comparative efficacy of single-coil versus dual-coil ICD leads: a meta-analysis of clinical outcomes.Future Cardiol. 2025 Mar;21(3):167-175. doi: 10.1080/14796678.2025.2459542. Epub 2025 Jan 30. Future Cardiol. 2025. PMID: 39882757
References
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical