Evaluation of bleb characteristics after implantation of the EX-PRESS™ glaucoma filtration device

Affiliations

Evaluation of bleb characteristics after implantation of the EX-PRESS™ glaucoma filtration device

Ashley E Laing et al. Mol Vis. 2012.

Abstract

Purpose: To compare bleb survival and histology after implantation of the EX-PRESS™ glaucoma filtration device versus silicone tubes in a rabbit model of filtration surgery.

Methods: Glaucoma filtration surgery was performed on one eye each of twelve New Zealand white rabbits. Eyes were randomized to implantation with the EX-PRESS™ filtration device (n=6) or a silicone tube (n=6). Bleb vascularity was evaluated at three and six weeks using a standard scale. At 6 weeks, eyes were enucleated and a histologic analysis was performed. Bleb survival was also recorded for the two groups.

Results: Histologically, a thin capsule consisting of mild fibroblast proliferation associated with intercellular collagen was present around both implants. Both groups demonstrated a mild infiltration of plasma cells and polymorphonuclear leukocytes. Bleb vascularity was similar between both groups at three and six weeks post-operatively. Bleb survival between the two groups was not significantly different.

Conclusions: Similar outcomes were noted after glaucoma filtration surgery using either silicone tubes or the EX-PRESS™ glaucoma filtration device in this rabbit model. Both implants appear to be relatively inert with little difference in biocompatibility and bleb survival.

PubMed Disclaimer

Similar articles

Cited by

References

    1. Quigley HA, Broman AT. The number of people with glaucoma worldwide in 2010 and 2020. Br J Ophthalmol. 2006;90:262–7. - PMC - PubMed
    1. Hendrick AM, Kahook MY. Ex-PRESS mini glaucoma shunt: surgical technique and review of clinical experience. Expert Rev Med Devices. 2008;5:673–7. - PubMed
    1. Maris PJ, Jr, Ishida K, Netland PA. Comparison of trabeculectomy with Ex-PRESS miniature glaucoma device implanted under scleral flap. J Glaucoma. 2007;16:14–9. - PubMed
    1. Good TJ, Kahook MY. Assessment of bleb morphologic features and postoperative outcomes after Ex-PRESS drainage device implantation versus trabeculectomy. Am J Ophthalmol. 2011;151:507–13. - PubMed
    1. Nyska A, Glovinsky Y, Belkin M, Epstein Y. Biocompatibility of the Ex-PRESS miniature glaucoma drainage implant. J Glaucoma. 2003;12:275–80. - PubMed

Publication types

LinkOut - more resources