Microwave ablation of ex vivo human liver and colorectal liver metastases with a novel 14.5 GHz generator
- PMID: 22235784
- DOI: 10.3109/02656736.2011.610428
Microwave ablation of ex vivo human liver and colorectal liver metastases with a novel 14.5 GHz generator
Abstract
Purpose: This study assessed the relationship between time, power and ablation size using a novel high-frequency 14.5 GHz microwave applicator in ex vivo human hepatic parenchyma and colorectal liver metastases. Previous examination has demonstrated structurally normal but non-viable cells within the ablation zone. This study aimed to further investigate how ablation affects these cells, and to confirm non-viability.
Materials and methods: Ablations were performed in ex vivo human hepatic parenchyma and tumour for a variety of time (10-180 s) and power (10-50 W) settings. Histological examination was performed to assess cellular anatomy, whilst enzyme histochemistry was used to confirm cellular non-viability. Transmission electron microscopy was used to investigate the subcellular structural effects of ablation within these fixed cells. Preliminary proteomic analysis was also performed to explore the mechanism of microwave cell death.
Results: Increasing time and power settings led to a predictable and reproducible increase in size of ablation. At 50 W and 180 s application, a maximum ablation diameter of 38.8 mm (±1.3) was produced. Ablations were produced rapidly, and at all time and power settings ablations remained spherical (longest:shortest diameter <1.2). Routine histological analysis using haematoxylin-eosin (H&E) confirmed well preserved cellular anatomy despite ablation. Transmission electron microscopy demonstrated marked subcellular damage. Enzyme histochemistry showed complete absence of viability in ablated tissue.
Conclusions: Large spherical ablation zones can be rapidly and reproducibly achieved in ex vivo human hepatic parenchyma and colorectal liver metastases using a 14.5 GHz microwave generator. Despite well preserved cellular appearance, ablated tissue is non-viable.
Similar articles
-
Microwave ablation: results with a 2.45-GHz applicator in ex vivo bovine and in vivo porcine liver.Radiology. 2006 Apr;239(1):94-102. doi: 10.1148/radiol.2383050262. Epub 2006 Feb 16. Radiology. 2006. PMID: 16484351
-
Microwave ablation of primary and secondary liver tumours: ex vivo, in vivo, and clinical characterisation.Int J Hyperthermia. 2017 Feb;33(1):34-42. doi: 10.1080/02656736.2016.1196830. Epub 2016 Jul 25. Int J Hyperthermia. 2017. PMID: 27443519 Clinical Trial.
-
A comparison of microwave ablation and bipolar radiofrequency ablation both with an internally cooled probe: results in ex vivo and in vivo porcine livers.Eur J Radiol. 2011 Jul;79(1):124-30. doi: 10.1016/j.ejrad.2009.12.009. Epub 2010 Jan 4. Eur J Radiol. 2011. PMID: 20047812
-
The role of microwave ablation in the management of hepatic colorectal metastases.Surgeon. 2011 Feb;9(1):33-7. doi: 10.1016/j.surge.2010.07.009. Epub 2010 Aug 21. Surgeon. 2011. PMID: 21195329 Review.
-
[Efficacy of local ablation therapy for liver metastasis from colorectal cancer--radiofrequency ablation and microwave coagulation therapy].Nihon Geka Gakkai Zasshi. 2001 May;102(5):390-7. Nihon Geka Gakkai Zasshi. 2001. PMID: 11394002 Review. Japanese.
Cited by
-
Evolution of surgical microwave ablation for the treatment of colorectal cancer liver metastasis: review of the literature and a single centre experience.Surg Today. 2015 Apr;45(4):407-15. doi: 10.1007/s00595-014-0879-3. Epub 2014 Mar 16. Surg Today. 2015. PMID: 24633931 Review.
-
Computational modelling of microwave tumour ablations.Int J Hyperthermia. 2013 Jun;29(4):308-17. doi: 10.3109/02656736.2013.799295. Int J Hyperthermia. 2013. PMID: 23738698 Free PMC article. Review.
-
Percutaneous microwave ablation applications for liver tumors: recommendations for COVID-19 patients.Heliyon. 2021 Mar 6;7(3):e06454. doi: 10.1016/j.heliyon.2021.e06454. eCollection 2021 Mar. Heliyon. 2021. PMID: 33748501 Free PMC article. Review.
-
Dynamic imaging and pathological changes in pig liver after MR-guided microwave ablation.BMC Cancer. 2018 Apr 6;18(1):397. doi: 10.1186/s12885-018-4157-4. BMC Cancer. 2018. PMID: 29625559 Free PMC article.
-
Use of microwave ablation for thermal treatment of solid tumors with different shapes and sizes-A computational approach.PLoS One. 2020 Jun 15;15(6):e0233219. doi: 10.1371/journal.pone.0233219. eCollection 2020. PLoS One. 2020. PMID: 32542034 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical