Noninvasive LV pressure estimation using subharmonic emissions from microbubbles
- PMID: 22239898
- PMCID: PMC6054451
- DOI: 10.1016/j.jcmg.2011.08.017
Noninvasive LV pressure estimation using subharmonic emissions from microbubbles
Abstract
To develop a new noninvasive approach to quantify left ventricular (LV) pressures using subharmonic emissions from microbubbles, an ultrasound scanner was used in pulse inversion grayscale mode; unprocessed radiofrequency data were obtained with pulsed wave Doppler from the aorta and/or LV during Sonazoid infusion. Subharmonic data (in dB) were extracted and processed. Calibration factor (mm Hg/dB) from the aortic pressure was used to estimate LV pressures. Errors ranged from 0.19 to 2.50 mm Hg when estimating pressures using the aortic calibration factor, and were higher (0.64 to 8.98 mm Hg) using a mean aortic calibration factor. Subharmonic emissions from ultrasound contrast agents have the potential to noninvasively monitor LV pressures.
Copyright © 2012 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.
Figures






References
-
- Chatterjee K. The Swan-Ganz catheters: past, present and future: a viewpoint. Circulation. 2009;119:147–52. - PubMed
-
- Solomon S, Stevenson L. Recalibrating the barometer: is it time to take a critical look at noninvasive approaches to measuring filling pressures? Circulation. 2009;119:13–5. - PubMed
-
- Shi W, Forsberg F, Raichlen J, Needleman L, Goldberg B. Pressure dependence of subharmonic signals from contrast microbubbles. Ultrasound Med Biol. 1999;25:275–83. - PubMed
-
- Forsberg F, Liu J, Shi W, Furuse J, Shimizu M, Goldberg B. In vivo pressure estimation using subharmonic contrast microbubble signals: proof of concept. IEEE Trans Ultrason, Ferroelec Freq Contr. 2005;52:581–3. - PubMed
Publication types
MeSH terms
Substances
Grants and funding
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical
Miscellaneous