Respiration-correlated image guidance is the most important radiotherapy motion management strategy for most lung cancer patients
Abstract
Purpose: The purpose of this study was to quantify the effects of four-dimensional computed tomography (4DCT), 4D image guidance (4D-IG), and beam gating on calculated treatment field margins in a lung cancer patient population.
Materials and methods: Images were acquired from 46 lung cancer patients participating in four separate protocols at three institutions in Europe and the United States. Seven patients were imaged using fluoroscopy, and 39 patients were imaged using 4DCT. The magnitude of respiratory tumor motion was measured. The required treatment field margins were calculated using a statistical recipe (van Herk M, et al. Int J Radiat Oncol Biol Phys 2000;474:1121-1135), with magnitudes of all uncertainties, except respiratory peak-to-peak displacement, the same for all patients, taken from literature. Required margins for respiratory motion management were calculated using the residual respiratory tumor motion for each patient for various motion management strategies. Margin reductions for respiration management were calculated using 4DCT, 4D-IG, and gated beam delivery.
Results: The median tumor motion magnitude was 4.4 mm for the 46 patients (range 0-29.3 mm). This value corresponded to required treatment field margins of 13.7 to 36.3 mm (median 14.4 mm). The use of 4DCT, 4D-IG, and beam gating required margins that were reduced by 0 to 13.9 mm (median 0.5 mm), 3 to 5.2 mm (median 5.1 mm), and 0 to 7 mm (median 0.2 mm), respectively, to a total of 8.5 to 12.4 mm (median 8.6 mm).
Conclusion: A respiratory management strategy for lung cancer radiotherapy including planning on 4DCT scans and daily image guidance provides a potential reduction of 37% to 47% in treatment field margins. The 4D image guidance strategy was the most effective strategy for >85% of the patients.
Copyright © 2012 Elsevier Inc. All rights reserved.
Similar articles
-
Dynamic volume vs respiratory correlated 4DCT for motion assessment in radiation therapy simulation.Med Phys. 2012 May;39(5):2669-81. doi: 10.1118/1.4704498. Med Phys. 2012. PMID: 22559637
-
Craniocaudal safety margin calculation based on interfractional changes in tumor motion in lung SBRT assessed with an EPID in cine mode.Int J Radiat Oncol Biol Phys. 2012 Jul 1;83(3):1064-9. doi: 10.1016/j.ijrobp.2011.07.043. Epub 2012 Jan 13. Int J Radiat Oncol Biol Phys. 2012. PMID: 22245190
-
Comparison of the planning target volume based on three-dimensional CT and four-dimensional CT images of non-small-cell lung cancer.Radiother Oncol. 2011 May;99(2):176-80. doi: 10.1016/j.radonc.2011.03.015. Epub 2011 May 4. Radiother Oncol. 2011. PMID: 21549440
-
[Image-guided radiotherapy in lung cancer].Cancer Radiother. 2018 Oct;22(6-7):602-607. doi: 10.1016/j.canrad.2018.06.009. Epub 2018 Aug 10. Cancer Radiother. 2018. PMID: 30104150 Review. French.
-
Image-guided radiotherapy and motion management in lung cancer.Br J Radiol. 2015 Jul;88(1051):20150100. doi: 10.1259/bjr.20150100. Epub 2015 May 8. Br J Radiol. 2015. PMID: 25955231 Free PMC article. Review.
Cited by
-
Preliminary comparison of the registration effect of 4D-CBCT and 3D-CBCT in image-guided radiotherapy of Stage IA non-small-cell lung cancer.J Radiat Res. 2017 Nov 1;58(6):854-861. doi: 10.1093/jrr/rrx040. J Radiat Res. 2017. PMID: 28992047 Free PMC article.
-
Objected constrained registration and manifold learning: a new patient setup approach in image guided radiation therapy of thoracic cancer.Med Phys. 2013 Apr;40(4):041710. doi: 10.1118/1.4794489. Med Phys. 2013. PMID: 23556880 Free PMC article. Clinical Trial.
-
Evaluation of target volume and dose accuracy in intrafractional cases of lung cancer based on 4D-CT and 4D-CBCT images using an in-house dynamic thorax phantom.Rep Pract Oncol Radiother. 2022 May 19;27(2):360-370. doi: 10.5603/RPOR.a2022.0029. eCollection 2022. Rep Pract Oncol Radiother. 2022. PMID: 36299381 Free PMC article.
-
A hybrid 2D/4D-MRI methodology using simultaneous multislice imaging for radiotherapy guidance.Med Phys. 2022 Sep;49(9):6068-6081. doi: 10.1002/mp.15802. Epub 2022 Jun 22. Med Phys. 2022. PMID: 35694905 Free PMC article.
-
Variation in target volume and centroid position due to breath holding during four-dimensional computed tomography scanning: A phantom study.J Appl Clin Med Phys. 2020 Jan;21(1):11-17. doi: 10.1002/acm2.12692. Epub 2019 Aug 5. J Appl Clin Med Phys. 2020. PMID: 31385421 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical