. 2012 Jan-Feb;13(1):61-5.
doi: 10.3348/kjr.2012.13.1.61. Epub 2011 Dec 23.

A retrospective review on feasibility and safety of a new pneumatic compression device for femoral arteriotomy hemostasis

Affiliations

A retrospective review on feasibility and safety of a new pneumatic compression device for femoral arteriotomy hemostasis

Minho Kang et al. Korean J Radiol. 2012 Jan-Feb.

Abstract

Objective: To report our initial experience on the technical feasibility and safety for hemostasis of a new pneumatic compression device in patients undergoing femoral arteriotomy.

Materials and methods: This study included 40 consecutive patients in whom hemostasis after transfemoral catheterization was readered by using a pneumatic compression device consisting of an inflatable bulb-containing main body and four pieces of supplementary tape. Medical records were retrospectively reviewed for outcomes and complications of hemostasis. Technical success was defined as achieving immediate hemostasis 10 minutes after applying the device over the arteriotomy sites, and clinical success was defined as the ability to ambulate after 4 hours of bed rest without any complications.

Results: Technical and clinical success was achieved in 38 (95%) and 37 (93%) patients, respectively. In two patients, hemostasis was achieved after conversion to manual compression. One patient required sand bag placement after removal of the device to control minimal oozing of blood. No patients had late complications.

Conclusion: The new pneumatic compression device provides effective and safe hemostasis after transfemoral catheterization in selected patient populations.

Keywords: Chemoembolization; Femoral artery; Hemostasis; Punctures.

PubMed Disclaimer

Figures

Fig. 1
Fig. 1
GH-150 applied to model doll. Supplementary tapes (arrows) are applied over wings of main body.
Fig. 2
Fig. 2
Non-enhanced CT scan of 63-year-old male patient. Femoral artery (arrows) is firmly compressed by inflated GH-150 (arrowheads) which is considered to lead to weak pulsation of ipsilateral dorsalis pedis.

Similar articles

Cited by

References

    1. Madigan JB, Ratnam LA, Belli AM. Arterial closure devices. A review. J Cardiovasc Surg (Torino) 2007;48:607–624. - PubMed
    1. Bent CL, Kyriakides C, Matson M. Femoral artery stenosis following percutaneous closure using a starclose closure device. Cardiovasc Intervent Radiol. 2008;31:814–816. - PubMed
    1. Brueck M, Bandorski D, Rauber K, Boening A. Percutaneous transluminal dilatation of inadvertent partial or complete occlusion of the femoral artery caused by Angio-Seal deployment for puncture site closure after cardiac catheterization. J Invasive Cardiol. 2010;22:353–357. - PubMed
    1. Cooper CL, Miller A. Infectious complications related to the use of the angio-seal hemostatic puncture closure device. Catheter Cardiovasc Interv. 1999;48:301–303. - PubMed
    1. Jang JJ, Kim M, Gray B, Bacharach JM, Olin JW. Claudication secondary to Perclose use after percutaneous procedures. Catheter Cardiovasc Interv. 2006;67:687–695. - PubMed